1.Titel 1.077B_e.psd - Petrick-GmbH
1.Titel 1.077B_e.psd - Petrick-GmbH
1.Titel 1.077B_e.psd - Petrick-GmbH
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>Petrick</strong><br />
Technical Specification<br />
Small stationary anode x-ray tube P470/8.35/12GIB (<strong>1.077B</strong>)<br />
For customer disposal available<br />
with wire or case connections.<br />
1. Geometry: Drawing for assembling<br />
2. Tilt angle: Target angle 12°<br />
3. Inherent filtration: 0,76...1,04 mm Al-equivalent<br />
4. Type of operating: DC<br />
5. Focal spot: 0,4 IEC 60336 grid controlled<br />
73,0 +1,5<br />
-1,5<br />
35,0 +0,2<br />
-0,2<br />
with R G = 28,0 kΩ<br />
at I Rö = 7 mA (U G = 196 V)<br />
6. Operating point: U Rö = 70 kV, I Rö = 7 mA<br />
7. max.operating voltage: 70 kV<br />
8. max.rating: 7 mA<br />
9. Exposure time: 3 s<br />
U H = 5,20 V ± 3%, I H = 1,40 A ± 3%<br />
10. Pulse ratio: 1 exposure 60 s break<br />
11. Power continuesly: 50 W<br />
12. Anode cooling: Convection cooling<br />
13. Insulation medium: Oil<br />
14. Anode material: Tungsten<br />
x-ray<br />
Design changes in the sense of the development reservations.<br />
3,5<br />
Ø 1,5<br />
M 4<br />
Ø 10,0 +0,0<br />
-0,3<br />
Ø 38,0<br />
M 1:1<br />
Ø 41,0 +0,1<br />
-0,1<br />
<strong>GmbH</strong><br />
Grid controlled<br />
Tube Voltage DC<br />
Rg<br />
Filament Voltage AC<br />
grid 3<br />
x-ray<br />
3<br />
2<br />
1<br />
+<br />
-<br />
2 heating<br />
1 heating
P470/8.35/12 GIB<br />
Radiation protection<br />
-Emitter can emit harmful X-rays<br />
-There is a danger of implosion<br />
-User is account for personal protection by a housing<br />
<strong>Petrick</strong><br />
<strong>GmbH</strong><br />
Plastic<br />
Pb<br />
Mo<br />
Ni<br />
W<br />
Cu<br />
FeNiCo<br />
Glass