28.05.2015 Views

BioTector Series 4 TOC Analyzer Technical Specifications

BioTector Series 4 TOC Analyzer Technical Specifications

BioTector Series 4 TOC Analyzer Technical Specifications

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

BIOTECTOR ANALYTICAL SYSTEMS LTD<br />

SPECIFICATION SHEET<br />

ANALYSIS PARAMETERS<br />

Principal Operation: Patented Two-Stage Advanced Oxidation Process using<br />

Hydroxyl Radicals.<br />

PAGE 2<br />

Measured Components:<br />

<strong>TOC</strong> (NPOC)<br />

TIC<br />

TC<br />

VOC / POC<br />

True <strong>TOC</strong> (TC – TIC)<br />

After Correlation Tests:<br />

COD<br />

BOD<br />

Principal Operation: Batch Samples<br />

Frequency: <strong>TOC</strong> < 6.5 minutes<br />

Monitoring Ranges: Ultra Low Range 0-5000µg/l (ppb)<br />

Low Range 0-1mg/l to 0-10mg/l (ppm)<br />

Standard Range 0-10mg/l to 0-10,000mg/l<br />

High Range 0-15mg/l to 0-20,000mg/l<br />

Ultra High Range 0-20mg/l to 0-100,000mg/l<br />

Other measuring ranges are available upon request.<br />

Exceedence Tracking:<br />

Records Full Tracking to Maximum Range<br />

REPEATABILITY:<br />

DETECTION LIMIT:<br />

<strong>TOC</strong><br />

<strong>TOC</strong><br />

Ultra Low Range ± 3% or ± 10ppb of the reading 20ppb<br />

Low Range ± 3% or ± 0.2mgC/l of the reading 0.4mgC/l<br />

Standard Range ± 3% or ± 0.3mgC/l of the reading 0.6mgC/l<br />

High Range ± 3% or ± 30mgC/l of the reading 60mgC/l<br />

Ultra High Range ± 3% or ± 300mgC/l of the reading 600mgC/l<br />

Sample Volume: 0.02 - 8.0ml<br />

Particle Size:<br />

Up to 2 mm Ø, soft particulates<br />

Filtration:<br />

Not required<br />

Signal Drift:<br />

< 5% per year<br />

Ambient Temperature: 5 – 40°C<br />

Humidity:<br />

5 - 85%, non-condensing<br />

Chloride Interference:<br />

All Ranges<br />

<strong>TOC</strong><br />

None<br />

<strong>BioTector</strong> Analytical Systems Limited<br />

Raffeen House, Ringaskiddy,<br />

Co. Cork,<br />

Ireland<br />

Tel: +353-21-4374237<br />

Fax: +353-21-4374236<br />

e-mail: info@biotector.com<br />

www.biotector.com