National Type Evaluation Program - Mettler Toledo
National Type Evaluation Program - Mettler Toledo
National Type Evaluation Program - Mettler Toledo
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>National</strong> Conference on Weights and Measures<br />
15245 Shady Grove Road, Suite 130 • Rockville, MD 20850<br />
Certificate Number: 01-049<br />
Page 1 of 2<br />
<strong>National</strong> <strong>Type</strong> <strong>Evaluation</strong> <strong>Program</strong><br />
Certificate of Conformance<br />
for Weighing and Measuring Devices<br />
For:<br />
Bench/Portable Scale<br />
Load Cell Electronic<br />
Model: Champ Square Scale CQ Series<br />
n max : 5 000<br />
e min : 0.005 lb (0.002 kg)<br />
Capacity: 20 lb (10 kg) to 500 lb (250 kg)<br />
Platform: See below<br />
Accuracy Class: III<br />
Submitted by:<br />
Ohaus Corporation<br />
19A Chapin Road<br />
Pine Brook, NJ 07058-9777<br />
Phone: (973) 377-9000<br />
Fax: (973) 593-0359<br />
Contact: Robert Hansen<br />
Standard Features and Options<br />
Painted carbon steel or stainless steel construction.<br />
Platform size from: 12 in x 12 in (144 in 2 ) to 24 in x 24 in (576 in 2 ).<br />
Load cells used:<br />
Carbon steel version:<br />
<strong>Mettler</strong> <strong>Toledo</strong> Model MT1241 - 30, 50, 100, 150 kg capacity<br />
<strong>Mettler</strong> <strong>Toledo</strong> Model MT1260 - 500 kg capacity<br />
Stainless steel version: <strong>Mettler</strong>-<strong>Toledo</strong> Model SSP -30, 50, 100, 200, 500 kg capacity<br />
The Champ Square Scale Series consists of the Champ Square Platform (Certificate of Conformance xx-xxx)<br />
interfaced to one of the following approved Indicating Elements:<br />
Temperature Range: -10 o C to 40 o C (14 o F to 104 o F)<br />
Ohaus CD11 Certificate of Conformance 99-100<br />
Ohaus CD31 Certificate of Conformance 99-071<br />
Ohaus CD33 Certificate of Conformance 99-071<br />
Ohaus CW11 Certificate of Conformance 99-100<br />
This device was evaluated under the <strong>National</strong> <strong>Type</strong> <strong>Evaluation</strong> <strong>Program</strong> (NTEP) and was found to comply with the applicable technical<br />
requirements of Handbook 44, "Specifications, Tolerances, and Other Technical Requirements for Weighing and Measuring Devices."<br />
<strong>Evaluation</strong> results and device characteristics necessary for inspection and use in commerce are on the following pages.<br />
Effective Date: July 12, 2001<br />
Louis E. Straub<br />
Chairman, NCWM, Inc.<br />
G. Weston Diggs<br />
Chairman, <strong>National</strong> <strong>Type</strong> <strong>Evaluation</strong> <strong>Program</strong> Committee<br />
Issue date: July 12, 2001<br />
Note: The <strong>National</strong> Conference on Weights and Measures does not "approve","recommend", or "endorse" any proprietary product or material, either as a<br />
single item or as a class or group. Results shall not be used in advertising or sales promotion to indicate explicit or implicit endorsement of the product or<br />
material by the NCWM.
Certificate Number: 01-049<br />
Page 2 of 2<br />
Ohaus, Corporation<br />
Bench/Counter Scale<br />
Load Cell Electronic<br />
Model: Champ Square Scale CQ Series<br />
Application: For use in general purpose weighing applications.<br />
Identification: The required information is on a label glued to the frame of the scale under the platter.<br />
Sealing: The device is sealed at the indicating element as per the Certificate of Conformance for the particular<br />
indicating element used.<br />
Test Conditions: The Champ Square Scale consists of a Champ Square Platform interfaced to either Ohaus<br />
Indicator Models CD11, CD31,CD33, or CW11. The test conditions for the Champ Square Platform are<br />
repeated here for reference: The emphasis of the evaluation was on the device design, marking, operation and<br />
compliance with influence factor requirements. The Ohaus Champ (10 kg and 500 lb) weighing elements were<br />
interfaced with <strong>Mettler</strong>-<strong>Toledo</strong> Hawk (Certificate of Conformance number 99-054) indicators and submitted for<br />
evaluation. Several increasing/decreasing load tests and shift tests were performed. The devices were tested<br />
over a temperature range of -10 o C to 40 o C (14 o F to 104 o F). A load of approximately one-half capacity was<br />
applied to the bases over 100 000 times (each). The scale was tested periodically during this time.<br />
The results of these evaluations and a review of technical information supplied by the manufacturer indicate the<br />
device conforms to the applicable requirements of NIST Handbook 44<br />
<strong>Type</strong> <strong>Evaluation</strong> Criteria Used: NIST Handbook 44, 2001 Edition<br />
Tested By: W. West (OH), T. Lucas (OH)<br />
Information Reviewed By: S. Patoray (NCWM)