25.04.2015 Views

manual program jaminan mutu dalam perkhidmatan radioterapi

manual program jaminan mutu dalam perkhidmatan radioterapi

manual program jaminan mutu dalam perkhidmatan radioterapi

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

KEMENTERIAN KESIHATAN MALAYSIA<br />

MANUAL PELAKSANAAN<br />

PROGRAM JAMINAN MUTU (QAP)<br />

DALAM PERKHIDMATAN<br />

RADIOTERAPI<br />

Disediakan oleh:<br />

Kumpulan Kerja Pelaksanaan Program Jaminan Mutu Dalam Perkhidmatan<br />

Radioterapi Di Bawah Akta Perlesenan Tenaga Atom 1984 (Akta 304)<br />

Kementerian Kesihatan Malaysia<br />

Julai 2012


KEMENTERIAN KESIHATAN MALAYSIA<br />

1. PENGENALAN<br />

Program Jaminan Mutu (Quality Assurance Programme-QAP) di Kementerian<br />

Kesihatan Malaysia telah dilancarkan pada tahun 1985. Objektif pelaksanaan QAP<br />

adalah untuk memastikan pelanggan mendapat faedah daripada <strong>perkhidmatan</strong> yang<br />

disediakan pada tahap yang optima dengan sumber yang sedia ada. Indikator-indikator<br />

diwujudkan untuk memantau kualiti pelbagai <strong>perkhidmatan</strong> dari aspek penjagaan,<br />

pengurusan pelanggan, resources utilization dan kepuasan pelanggan.<br />

Dalam konteks pelaksanaan QAP yang menggunakan sinaran mengion bagi tujuan<br />

terapi, kualiti rawatan ke atas pesakit perlu diutamakan. Manual ini dijadikan sebagai<br />

satu rujukan untuk membantu pelaksanaan QAP <strong>dalam</strong> <strong>perkhidmatan</strong> <strong>radioterapi</strong>.<br />

Objektif <strong>program</strong> ini adalah untuk memastikan kualiti <strong>perkhidmatan</strong> yang diberikan oleh<br />

institusi-institusi perubatan kerajaan dan swasta mampu meningkatkan tahap<br />

keselamatan pesakit, pekerja dan orang awam daripada risiko yang mungkin timbul<br />

akibat penggunaan radas penyinaran dan bahan radioaktif. Fokus utama <strong>manual</strong> ini<br />

disediakan adalah untuk menjadi panduan kepada <strong>program</strong> peningkatan kualiti secara<br />

berterusan.<br />

Keperluan pelaksanaan QAP ini adalah memenuhi kehendak Peraturan 53(1) <strong>dalam</strong><br />

Peraturan-Peraturan Perlesenan Tenaga Atom (Perlindungan Sinaran Keselamatan<br />

Asas) 2010 yang mengkehendaki pemegang lesen mewujudkan suatu <strong>program</strong><br />

<strong>jaminan</strong> <strong>mutu</strong> yang komprehensif bagi dedahan perubatan dengan penglibatan pakar<br />

berkelayakan yang sesuai <strong>dalam</strong> bidang yang berkaitan sebagaimana yang dinyatakan<br />

oleh pihak berkuasa yang berkenaan.<br />

Selain itu, Peraturan 41(e) <strong>dalam</strong> Peraturan berkenaan juga mengkehendaki tiap-tiap<br />

pemegang lesen atau majikan memastikan bahawa bagi penggunaan sinaran secara<br />

terapeutik termasuk teleterapi atau brakiterapi, penentukuran, dosimetri dan <strong>program</strong><br />

<strong>jaminan</strong> <strong>mutu</strong> yang dinyatakan oleh pihak berkuasa yang berkenaan dijalankan oleh<br />

atau di bawah penyeliaan seorang pakar yang berkelayakan <strong>dalam</strong> fizik perubatan.<br />

Mesyuarat Jawatankuasa Penasihat Radiologi (RAC) ke-37 yang diadakan pada 15hb.<br />

Julai 2010 telah me<strong>mutu</strong>skan supaya pelaksanaan QAP <strong>dalam</strong> <strong>perkhidmatan</strong><br />

<strong>radioterapi</strong> di sektor kerajaan dan swasta diperkukuhkan. Melalui pelaksanaan QAP<br />

tersebut, pusat-pusat <strong>radioterapi</strong> akan dapat menerap dan mengamalkan budaya kualiti<br />

dan selamat <strong>dalam</strong> penggunaan sinaran mengion untuk tujuan perubatan kepada<br />

pesakit, pekerja dan orang awam.<br />

Untuk memastikan QAP dilaksanakan dengan berkesan, jabatan berkaitan perlu<br />

mewujudkan Jawatankuasa QAP yang terdiri daripada pakar onkologi, ahli fizik<br />

perubatan, juru x-ray terapi dan lain-lain profesion yang terlibat.<br />

1


KEMENTERIAN KESIHATAN MALAYSIA<br />

2. OBJEKTIF<br />

Objektif pelaksanaan QAP <strong>dalam</strong> <strong>perkhidmatan</strong> <strong>radioterapi</strong> adalah :<br />

2.1 Meningkatkan kualiti <strong>perkhidmatan</strong> <strong>radioterapi</strong><br />

2.2 Memastikan ketepatan dan kejituan <strong>dalam</strong> rawatan pesakit<br />

2.3 Memastikan penggunaan sumber yang sedia ada secara efektif dan selamat<br />

2.4 Memenuhi dan mematuhi keperluan peraturan-peraturan di bawah Akta<br />

Perlesenan Tenaga Atom 1984 (Akta 304)<br />

3. PELAKSANAAN QAP SEBAGAI KEPERLUAN REGULATORI<br />

QAP hendaklah dilaksanakan di setiap pusat <strong>radioterapi</strong> dengan merangkumi elemenelemen<br />

seperti di bawah:-<br />

3.1 Indikator - “The rate of discrepancies of treatment verification images”<br />

Indikator “The rate of discrepancies of treatment verification images” hendaklah<br />

dipantau dan dianalisa di setiap pusat <strong>radioterapi</strong>. Laporan analisa tersebut<br />

termasuk tindakan pembaikan yang telah diambil perlu dikaji setiap tahun.<br />

Perbandingan imej yang diambil semasa rawatan dan proses simulasi adalah<br />

bertujuan untuk mengesahkan ketepatan rawatan yang diberi, terutamanya<br />

melibatkan organ kritikal yang berdekatan dengan kawasan rawatan. Ia hendaklah<br />

dilakukan terhadap sekurang-kurangnya 50% pesakit yang menjalani rawatan<br />

radikal <strong>dalam</strong> masa 3 hari pertama di permulaan rawatan.<br />

Maklumat lanjut mengenai kaedah pelaksanaan indikator di Lampiran A.<br />

3.2 Kawalan kualiti (QC) bagi radas penyinaran <strong>radioterapi</strong> dan kemudahan<br />

berkaitan termasuk prosidur khas<br />

Semua radas penyinaran <strong>radioterapi</strong> dan kemudahan berkaitan termasuk prosidur<br />

khas hendaklah diselenggara dan dikalibrasi secara berkala. Prosidur khas adalah<br />

seperti Stereotactic Radiosurgery/Radiotherapy (SRS/SRT), Intensity Modulated<br />

Radiation Therapy (IMRT) dan Interstitial Brachytherapy.<br />

Kawalan <strong>mutu</strong> (QC) bagi setiap radas penyinaran <strong>radioterapi</strong> hendaklah memenuhi<br />

standard prestasi dan keselamatan yang ditetapkan seperti di Lampiran B.<br />

Laporan berkenaan hendaklah disahkan oleh Juruperunding Fizik Perubatan atau<br />

Ahli Fizik Perubatan Hospital yang diiktiraf oleh KKM.<br />

2


KEMENTERIAN KESIHATAN MALAYSIA<br />

3.3 Pendidikan Profesional Berterusan (CPE)<br />

Semua personel hendaklah menghadiri <strong>program</strong> CPE yang diiktiraf oleh pihak<br />

berkuasa di bawah Akta 304 untuk meningkatkan pengetahuan dan kompetensi.<br />

Tempoh latihan hendaklah sekurang-kurangnya 12 jam terkumpul setahun dengan<br />

sekurang-kurangnya merangkumi salah satu daripada topik-topik seperti berikut:-<br />

<br />

<br />

<br />

<br />

<br />

<br />

<br />

<br />

Clinical Aspects of Radiobiology<br />

Radiation Safety and Protection<br />

External Beam Radiation Dosimetry<br />

External Beam Radiation Therapy<br />

External Beam Treatment Planning<br />

Brachytherapy<br />

Quality Assurance Programme in Radiotherapy<br />

Legislation and Regulatory Requirement<br />

Bukti seperti salinan sijil penyertaan atau senarai kehadiran hendaklah disahkan<br />

oleh Ketua Jabatan atau agensi penganjur bagi setiap personel terlibat.<br />

4. PENGURUSAN REKOD<br />

Kesemua elemen QAP yang dinyatakan seperti di Perkara 3 hendaklah dikemukakan<br />

kepada KKM setiap tahun sebagai keperluan regulatori, iaitu Annual Analysis Report for<br />

the Rate of Discrepancies of Treatment Verification Images (Form B), sijil/laporan QC<br />

dan salinan sijil/bukti kehadiran CPE.<br />

Rekod-rekod berkenaan dan rekod-rekod yang berkaitan hendaklah diurus dan<br />

disimpan <strong>dalam</strong> tempoh yang ditetapkan oleh KKM.<br />

3


KEMENTERIAN KESIHATAN MALAYSIA<br />

DEFINISI<br />

Central axis<br />

Gating<br />

Image acquisition<br />

Image registration<br />

Intact<br />

Integrity<br />

Isocentre<br />

Set-up errors (field<br />

placement errors)<br />

Systematic errors<br />

The imaginary axis or line which passes through both the x-ray<br />

source and the isocentre of the machine.<br />

Gating or respiratory gating is a treatment delivery technique<br />

which allows the treatment of tumours at certain defined points<br />

in the respiratory cycle.<br />

The process of acquiring image data. In the context of<br />

geometric verification, it may be a 2D (planar) or a 3D (volume)<br />

set of data, and may be obtained with either ionizing or nonionising<br />

radiation.<br />

Methods of aligning two 3D image sets; for example, CT, MRI,<br />

PET etc. Image sets may be overlaid or structures may be<br />

mapped between the sets.<br />

Describes something that is complete and that has not been<br />

damaged in any way.<br />

The absence of unintended changes or errors.<br />

A single point within the treatment room (in space) towards<br />

which the radiation beam always points. The central beam axis<br />

passes through this point and, on a LINAC, the three principle<br />

rotational movements of gantry, collimator and floor are all<br />

around axes which intersect at this point. For a tomotherapy<br />

machine, it is a point of intersection between the centre of the<br />

scan plane and the axis of rotation of the scan cycle.<br />

Any geometric displacement in patient set-up (localisation) with<br />

respect to the desired reference defined by the treatment plan,<br />

which is present at the time of patient set-up during delivery.<br />

Geometric displacements in patient set-up (localisation) with<br />

respect to the desired reference defined by the treatment plan,<br />

which are similar in both magnitude and direction for each<br />

treatment fraction. These are primarily due to systematic<br />

differences in equipment or protocol throughout the<br />

radiotherapy process (that is, from pre-treatment imaging to<br />

treatment planning to pre-treatment verification etc.).<br />

Tolerances The permitted observed variation in a parameter or<br />

measurement from its desired value.<br />

4


KEMENTERIAN KESIHATAN MALAYSIA<br />

SINGKATAN<br />

AAPM American Association of Physicists in Medicine<br />

AEC Automatic Exposure Control<br />

AQA Automatic Quality Assurance<br />

CT Computed Tomography<br />

CTDI Computed Tomography Dose Index<br />

deg Degree<br />

EPID Electronic Portal Imaging Device<br />

FWHM Full width at half maximum<br />

HU Hounsfield Unit<br />

HVL Half Value Layer<br />

IEC International Electro-technical Commission<br />

kV kilo voltage<br />

mMLC mini Multileaf Collimator<br />

MU Monitor Unit<br />

MV Megavoltage<br />

MVCT Megavoltage Computed Tomography<br />

PDD Percentage Depth Dose<br />

PDD 10 Percentage depth dose at water depth of 10 g/cm 2<br />

QA Quality Assurance<br />

QC Quality Control<br />

RMS Root Mean Square<br />

SAD Source to axis distance<br />

SDD Source to Diaphragm Distance<br />

TAR Tissue Air Ratio<br />

TMR Tissue Maximum Ratio<br />

TMR 20<br />

WF<br />

10<br />

Tissue Maximum Ratio in water at depths of 20 g/cm 2 and 10 g/cm 2<br />

Wedge Factor<br />

5


KEMENTERIAN KESIHATAN MALAYSIA<br />

RUJUKAN<br />

i. Akta Perlesenan Tenaga Atom 1984 (Akta 304).<br />

ii.<br />

Peraturan-Peraturan Perlesenan Tenaga Atom (Perlindungan Sinaran Keselamatan<br />

Asas) 2010.<br />

iii. Peraturan-Peraturan Perlindungan Sinaran (Perlesenan) 1986.<br />

iv. Ministry of Health Malaysia. Quality Assurance-A Problem Solving Approach. Kuala<br />

Lumpur: MOH.<br />

v. International Atomic Energy Agency. 2005. Radiation oncology physics: A handbook for<br />

teachers and students. Austria, Vienna: IAEA.<br />

vi. Kutcher GJ, Cola L, Gillin M, Hanson WF, Leibel S, Morton RJ and et al.<br />

Comprehensive QA for Radiation Oncology: Report of AAPM Radiation Therapy<br />

Committee Task Group 40. Medical Physics. 1994: 21(4).<br />

vii. Mutic S, Palta JR, Butker EK, Das IJ, Huq MS and et al. Quality assurance for<br />

computed-tomography simulators and the computed tomography-simulation process:<br />

Report of the AAPM Radiation Therapy Committee Task Group No. 66. Medical<br />

Physics. 2003: 30(10).<br />

viii. Klein EE, Hanley J, Bayouth J, Yin FF and et al. Quality assurance of medical<br />

accelerators: Task Group 142 report: Quality assurance of medical accelerators.<br />

Medical Physics. 2009: 36(9).<br />

ix. Dieterich S, Cavedon C, Chuang CF, Cohen AB and et al. Quality assurance for robotic<br />

radiosurgery: Report of AAPM TG 135. Medical Physics. 2011: 38(6).<br />

x. Langen KM, Papanikolaou N, Balog J, Crilly R, Followill D, Goddu AM and et al. QA for<br />

Helical Tomotherapy: Report of the AAPM Task Group 148. Medical Physics. 2010:<br />

37(9).<br />

xi. Khan FM. The Physics of Radiation Therapy. USA: Williams and Wilkins. 2003.<br />

6


LAMPIRAN A<br />

KEMENTERIAN KESIHATAN MALAYSIA<br />

STRUCTURE OF A PERFORMANCE INDICATOR<br />

FOR RADIOTHERAPY<br />

PROGRAMME<br />

AREA OF CONCERN<br />

INDICATOR<br />

: Radiotherapy Services<br />

: Performance Quality and Radiation Safety of Radiotherapy<br />

Services<br />

: The Rate Of Discrepancies of Treatment Verification Images<br />

DEFINITION OF TERM<br />

Treatment verification images are defined as the images taken on the treatment machine to<br />

confirm the accuracy of the intended treatment fields. It must be done within the first 3 days<br />

of starting treatment.<br />

RATIONALE<br />

The process of delivery of radiation is laborious, particularly for radical treatment. This is<br />

especially true for cases where important vital structures and organs are situated in the<br />

vicinity of the radiation portals.<br />

Simulating the treatment field portals is a process done to ensure that the radiation fields are<br />

where it is intended to be. However due to many technical factors such as setup difficulties,<br />

machine accuracy, transfer of data etc, this may not be reproducible at the time of treatment.<br />

Verification images done during the actual treatment can be compared to the simulation<br />

images and any discrepancies can be corrected before treatment is allowed to progress. This<br />

will help to maximize accuracy of treatment and will address systematic errors.<br />

TYPE OF INDICATOR<br />

NUMERATOR<br />

DENOMINATOR<br />

: This is a rate-based indicator.<br />

It is a measure of clinical, physics and maintenance services.<br />

: Total number of rejected verification images<br />

: Total number of verification images taken<br />

CALCULATION OF RATE : (Numerator/Denominator) x 100 %<br />

ACCEPTABLE<br />

DEVIATION<br />

: Head and neck – less than or equal to 3mm<br />

Extremities/abdomen/pelvis/thorax – less than or equal to<br />

5mm<br />

STANDARD : Discrepancy rate should be less than 25%<br />

7


KEMENTERIAN KESIHATAN MALAYSIA<br />

METHODOLOGY<br />

1. Data Collection<br />

i. Select patients for radical radiotherapy;<br />

ii. Acquire first uncorrected verification image within the first 3 days of radiotherapy;<br />

iii. Record in Form A.<br />

2. Data Analysis<br />

This involves transferring data from Form A to Form B. The tabulated data can then be<br />

analyzed to determine the types of errors. Data analysis can be carried out monthly, 3-6<br />

monthly or annually depending on the respective centre.<br />

3. Results<br />

The final presentation by using Form B shall be sent to the appropriate authority annually.<br />

4. Investigation protocol for abnormal discrepancies from intended treatment field<br />

i. Identify cause of discrepancies - from the table (Form A). The reason for the<br />

discrepancies will most likely be obvious from the analysis and presentation table.<br />

ii. Remedial action is carried out based on the cause ascertained.<br />

8


Form A<br />

KEMENTERIAN KESIHATAN MALAYSIA<br />

PORTAL VERIFICATION FORM<br />

MONTH: ...................<br />

No.<br />

Patient<br />

IC No.<br />

Start<br />

treatment<br />

Date<br />

Acquisition<br />

Method of<br />

image<br />

acquisition<br />

Treatment<br />

site<br />

Set up<br />

errors<br />

(mm)<br />

Accepted (/)/<br />

Rejected (X)<br />

*Type of<br />

verification<br />

errors<br />

Remarks<br />

1<br />

2<br />

3<br />

4<br />

5<br />

6<br />

7<br />

8<br />

9<br />

10<br />

* Type of verification errors:<br />

A: Isocentre Shift<br />

B: Incorrect Field Orientation (X,Y)<br />

C: Incorrect Field Size<br />

D: Shielding<br />

E: Others (Please specify)<br />

Prepared by : ...............................<br />

(Signature & Cop)<br />

9<br />

Date : .............................


Form B<br />

KEMENTERIAN KESIHATAN MALAYSIA<br />

ANNUAL ANALYSIS REPORT FOR<br />

THE RATE OF DISCREPANCIES OF TREATMENT VERIFICATION IMAGES<br />

Name of Hospital : ..................................................................<br />

Year : ..................................................................<br />

Month<br />

Jan<br />

Total number of<br />

verification images taken<br />

Total number of<br />

rejected verification<br />

images<br />

*Rate of discrepancies<br />

(%)<br />

Feb<br />

Mac<br />

Apr<br />

May<br />

Jun<br />

Jul<br />

Aug<br />

Sept<br />

Oct<br />

Nov<br />

Dec<br />

Total<br />

Total number of rejectedimages<br />

* The rate of discrepancies (%) <br />

100%<br />

Total number of verification images being taken<br />

Total number of patients per year undergoing radical treatment : ...........................................<br />

Total number of the above patients per year having verification done : ...........................................<br />

Percentage (%) of the above patients having verification done : ..........................................<br />

SHORTFALLS IN QUALITY<br />

Causes<br />

: …………………………………………………………………………………………….<br />

…………………………………………………………………………………………….<br />

…………………………………………………………………………………………….<br />

Corrective actions taken : …………………………………………………………………………………………….<br />

…………………………………………………………………………………………….<br />

…………………………………………………………………………………………….<br />

Verified by<br />

: …………………………………………………………………………………………….<br />

(Signature & Chop)<br />

10


LAMPIRAN B<br />

KEMENTERIAN KESIHATAN MALAYSIA<br />

STANDARD DAN KRITERIA PRESTASI DAN KESELAMATAN BAGI RADAS<br />

PENYINARAN RADIOTERAPI DAN KEMUDAHAN BERKAITAN<br />

Jadual 1<br />

Jadual 2<br />

Jadual 3<br />

Jadual 4<br />

Jadual 4.a<br />

Jadual 4.b<br />

Jadual 4.b.i<br />

: Simulator<br />

: CT Simulator<br />

: Irradiating Apparatus below 1 MV<br />

: Linear Accelerator (LINAC)<br />

: Additional QC Parameters for LINAC-Based Stereotactic<br />

Radiosurgery/Radiotherapy (SRS/SRT)<br />

: Additional QC Parameters for LINAC-Based Intensity<br />

Modulated Radiation Therapy (IMRT)<br />

: Additional QC Parameters for Multi-leaf Collimation<br />

(with differentiation of IMRT vs. non-IMRT machines)<br />

Jadual 4.b.ii : Additional QC Parameters for Dynamic/Universal/Virtual<br />

Wedges<br />

Jadual 5<br />

Jadual 6<br />

: Helical Tomotherapy<br />

: Cyberknife<br />

Jadual 7<br />

Jadual 7.a<br />

Jadual 7.b<br />

:<br />

:<br />

:<br />

Brachytherapy<br />

Remote Afterloading Brachytherapy Unit<br />

Special Procedures for Permanent Interstitial Brachytherapy<br />

Jadual 8<br />

Jadual 9<br />

Jadual 10<br />

: Treatment Planning System for Teletherapy<br />

: Imaging Systems (Portal Imaging and Image Guided Radiation<br />

Therapy)<br />

: Intraoperative Radiation Therapy (Low kV Photon)<br />

11


KEMENTERIAN KESIHATAN MALAYSIA<br />

Jadual 1: Simulator<br />

No. Parameters Tolerance Frequency<br />

1. Mechanical Checks<br />

i. Localising lasers 2 mm Daily<br />

ii. Optical Distance Indicator (ODI)<br />

2 mm<br />

iii. Field size indicator<br />

2 mm<br />

Radiation Safety Checks<br />

i. Radiation on/off warning light Functional Daily<br />

ii. Emergency button Functional Quarterly<br />

2. Mechanical Checks<br />

i. Gantry/collimator angle indicators 1º Monthly<br />

ii. Cross-hair centring<br />

2 mm diameter<br />

iii. Focal spot-axis indicator<br />

2 mm<br />

iv. Emergency/collision avoidance<br />

Functional<br />

v. Light/radiation field coincidence 2 mm or 1%<br />

vi. Collimator rotation isocentre<br />

2 mm diameter<br />

vii. Gantry rotation isocentre<br />

2 mm diameter<br />

viii. Couch rotation isocentre<br />

2 mm diameter<br />

ix. Coincidence of collimator, gantry,<br />

2 mm diameter<br />

couch axes and isocentre<br />

x. Table top sag 2 mm<br />

xi. Vertical travel of couch<br />

2 mm<br />

3. Radiographic and Fluoroscopic Checks<br />

i. Exposure reproducibility 10% Annually or<br />

ii. Exposure linearity 10%<br />

iii. kVp accuracy<br />

5% or 5kV whichever is<br />

greater<br />

iv. High and low contrast resolution<br />

Baseline<br />

v. Image Receptor Input Dose Rate Manufacturer’s Specification<br />

vi. Entrance Surface Dose Rate Limit<br />

a. Equipment with AEC<br />

b. Equipment without AEC<br />

vii. Leakage radiation<br />


KEMENTERIAN KESIHATAN MALAYSIA<br />

* Notes:<br />

i. The tolerances for simulator are stricter than for linear accelerators because errors at<br />

treatment planning stages on the simulator carry through to each treatment on the linear<br />

accelerator.<br />

ii.<br />

The maintenance services refer to those services that involved the changing of major<br />

parts of the machine, which might affect the dose given to the patient and the quality of<br />

the images produced.<br />

iii. The Quality Assurance (QA) <strong>program</strong> should be based on a thorough investigation for<br />

baseline standards at the time of the acceptance and commissioning of the equipment<br />

for clinical use. The procedures for acceptance tests should be followed to verify the<br />

manufacturer’s specifications and to establish baseline performance values for new or<br />

refurbished equipment, or for equipment following major repair. Once a baseline<br />

standard has been established, a protocol for periodic QA tests should be developed for<br />

the purpose of monitoring the reference performance values.<br />

13


KEMENTERIAN KESIHATAN MALAYSIA<br />

Jadual 2: CT Simulator<br />

No. Parameters Tolerance Frequency<br />

1. Mechanical Checks<br />

Localising lasers 2mm Daily<br />

Radiation Safety Checks<br />

i. Radiation on/off warning light Functional Daily<br />

ii. Emergency button Functional Quarterly<br />

2. X-Ray Generator<br />

i. Accuracy of kVp ± 5% or 5kV (whichever is greater) Annually<br />

ii. Accuracy of exposure time ± 10%<br />

iii. Exposure linearity (mR/mAs) Coefficient of linearity ≤ 0.1<br />

iv. Radiation output reproducibility Coefficient of variation ≤ 0.1<br />

3. Radiation Dosimetry<br />

i. Patient dosimetry (CTDI) Manufacturer specifications<br />

ii. Scout localisation image<br />

±20% of nominal value<br />

Annually<br />

4. Scan Localisation<br />

i. Axial scan localisation light<br />

accuracy<br />

ii. Isocentre alignment, sagittal and<br />

coronal localisation light accuracy<br />

±2 mm Annually<br />

±5 mm<br />

iii. Gantry tilt accuracy ±3º<br />

iv. Table index<br />

±0.5 mm<br />

v. Table position ±2.0 mm<br />

vi. Image scan width (sensitivity<br />

profile)<br />

a. Single slice ±0.5 mm (


KEMENTERIAN KESIHATAN MALAYSIA<br />

No. Parameters Tolerance Frequency<br />

ii. Hard copy display<br />

5% and 95% patches must be<br />

visible, no noticeable artefacts;<br />

Geometric distortion not exceed<br />

±1 mm;<br />

Optical density values must be<br />

within specified range.<br />

Annually<br />

6. Image Quality<br />

i. CT number uniformity ±5 HU Monthly<br />

ii. Image artefacts (transaxial and<br />

scan localisation images<br />

iii. Noise<br />

No significant artefacts<br />

Standard deviation of CT numbers<br />

varies as reciprocal square root of<br />

mAs<br />

Semi-<br />

Annually<br />

iv. Low contrast resolution 5 mm Annually<br />

v. High contrast resolution 1 mm holes (5lp/cm) Annually<br />

7. Quantitative Accuracy<br />

i. CT number constancy Value and standard deviation for Daily<br />

water remains relatively constant<br />

ii. CT number calibration Water : 0 ± 5HU Monthly<br />

iii. Accuracy of distance<br />

±1 mm Annually<br />

measurements (transaxial and<br />

scan localisation images)<br />

iv. CT number dependence on scan<br />

± 3HU<br />

thickness<br />

v. CT number dependence on<br />

phantom size<br />

± 20HU<br />

vi. CT number dependence on<br />

phantom position<br />

vii. CT number dependence on<br />

reconstruction algorithm<br />

± 5HU<br />

± 3HU<br />

8. Scattered and Leakage Radiation<br />

i. Scattered radiation


KEMENTERIAN KESIHATAN MALAYSIA<br />

No. Parameters Tolerance Frequency<br />

9. External Localisation Laser<br />

Orientation of gantry lasers with respect<br />

to the imaging plane<br />

±2 mm over the length of laser<br />

projection<br />

Monthly<br />

Spacing of lateral wall lasers with<br />

respect to lateral gantry lasers and scan<br />

plane<br />

±2 mm<br />

Orientation of wall lasers with respect to<br />

the imaging plane<br />

±2 mm over the length of laser<br />

projection<br />

Orientation of the ceiling laser with<br />

respect to the imaging plane<br />

±2 mm over the length of laser<br />

projection<br />

Orientation of the CT-scanner tabletop<br />

with respect to the imaging plane<br />

±2 mm over the length and width<br />

of the tabletop<br />

Table vertical and longitudinal motion<br />

±1 mm over the range of table<br />

16


KEMENTERIAN KESIHATAN MALAYSIA<br />

Jadual 3: Irradiating Apparatus below 1 MV<br />

No. Parameters Tolerance Frequency<br />

1. Mechanical Checks<br />

i. Mechanical integrity of the unit Functional Quarterly<br />

ii. Firmness of the locking devices<br />

Functional<br />

2. Radiation Safety Checks<br />

i. Door interlock Functional Daily<br />

ii. Radiation on/off warning light Functional Daily<br />

iii. Emergency button Functional Quarterly<br />

iv. Locking devices Functional Quarterly<br />

3. Dosimetry<br />

i. Absolute dose<br />

Semi-Annually<br />

ii. HVL test<br />

Standard recognised by Semi-Annually<br />

iii. Beam homogeneity film test<br />

the appropriate authority Semi-Annually<br />

iv. Cut out lead transmission factors<br />

Every new batch<br />

of lead<br />

v. Light/radiation field coincidence ±2 mm or 2% Semi-Annually<br />

4. i. Tube head leakage<br />

a) Contact therapy<br />

b) < 500 kVp<br />

1 mGy per hour at 5 cm<br />

from the tube housing<br />

10 mGy per hour at 1 m<br />

from the source<br />

Tube head<br />

change or<br />

service<br />

c) 500 kVp - 999 kVp<br />

0.1 percent of useful<br />

beam or 10 mGy per<br />

hour, whichever is greater<br />

at 1 m from the source<br />

ii. Applicator geometrical accuracy and<br />

beam axis indication on rotation<br />

iii. Dose monitor reproducibility<br />

iv. Dose monitor linearity<br />

v. Output consistency<br />

< 2 mm Quarterly<br />

< 0.5%<br />

2%<br />

3%<br />

Quarterly<br />

Quarterly<br />

Quarterly<br />

17


KEMENTERIAN KESIHATAN MALAYSIA<br />

Jadual 4: Linear Accelerator<br />

No. Parameters Tolerance Frequency<br />

1. Mechanical Checks<br />

i. Localising lasers 2 mm Daily<br />

ii. Optical Distance Indicator (ODI)<br />

2 mm<br />

iii. Field light<br />

Functional<br />

Safety Checks<br />

i. Door Interlock Functional<br />

ii. Audiovisual monitor<br />

Functional<br />

2. Dosimetry<br />

i. X-ray output constancy 2% Weekly<br />

ii. Electron output constancy 2%<br />

iii. X-ray beam flatness constancy 2% Quarterly<br />

iv. Electron beam flatness constancy 3%<br />

v. X-ray and electron symmetry 3%<br />

vi. Backup monitor constancy 2%<br />

3. Mechanical Checks<br />

i. Light/radiation field coincidence 2 mm for field up to 20 cm or<br />

± 1% for field over 20 cm<br />

ii. Gantry/collimator angle indicators<br />

1 deg<br />

iii. Wedge position<br />

2 mm (or 2% change in<br />

transmission factor)<br />

iv. Tray position<br />

2 mm<br />

v. Applicator position 2 mm<br />

vi. Field size indicators<br />

2 mm<br />

vii. Cross-hair centring<br />

2 mm diameter<br />

viii. Treatment couch position indicators<br />

2mm/1 deg<br />

ix. Latching of wedges, blocking tray<br />

Functional<br />

x. Jaw symmetry 2 mm<br />

Monthly<br />

Safety Checks<br />

i. Emergency off switches Functional<br />

ii. Wedge, electron cone interlocks<br />

Functional<br />

4. Dosimetry<br />

i. X-ray/electron output calibration<br />

constancy<br />

ii. Photon central axis parameter<br />

constancy (PDD, TAR)<br />

2% Annually<br />

2%<br />

18


KEMENTERIAN KESIHATAN MALAYSIA<br />

No. Parameters Tolerance Frequency<br />

iii. Electron central axis dosimetry<br />

parameter constancy (PDD)<br />

2 mm at therapeutic depth<br />

iv. Field size dependence of x-ray output<br />

constancy<br />

v. Output factor constancy for electron<br />

applicators<br />

2%<br />

2%<br />

vi. X-ray output constancy vs. gantry angle 2%<br />

vii. Electron output constancy vs. gantry<br />

2%<br />

angle<br />

viii. Wedge transmission factor constancy 2%<br />

ix. Monitor chamber linearity 1%<br />

x. Transmission factor constancy for all<br />

2%<br />

treatment accessories<br />

xi. Arc mode<br />

Manufacturers’ specification<br />

Safety Interlocks<br />

Follow manufacturers test procedures<br />

Functional<br />

Mechanical Checks<br />

i. Collimator rotation isocentre 2 mm diameter<br />

ii. Gantry rotation isocentre<br />

2 mm diameter<br />

iii. Couch rotation isocentre<br />

2 mm diameter<br />

iv. Coincidence of collimator, gantry, couch 2 mm diameter<br />

axes with isocentre<br />

v. Coincidence of radiation and<br />

2 mm diameter<br />

mechanical isocentre<br />

vi. Table top sag<br />

2 mm<br />

vii. Vertical travel of table<br />

2 mm<br />

Scattered radiation<br />

< 0.1mGy (10mR) per week<br />

Note:<br />

Absolute dose calibration needs to be performed after each maintenance service that involved the<br />

change of major parts which affect the dosimetry.<br />

19


KEMENTERIAN KESIHATAN MALAYSIA<br />

Jadual 4.a: Additional QC Parameters for LINAC-Based Stereotactic<br />

Radiosurgery/Radiotherapy (SRS/SRT)<br />

No. Parameters Tolerance Frequency<br />

1. Laser localisation check 1 mm Daily<br />

2. Winston Lutz test 1.5 mm Weekly (SRT)/<br />

Before<br />

treatment (SRS)<br />

3. Patient-specific plan verification<br />

i. Plan parameters check<br />

ii. Collision check in each field<br />

SRS/SRT field<br />

iii. MLC/mMLC/Cone accuracy with<br />

light field in each SRS/SRT field<br />

Pass/fail<br />

Before first<br />

treatment for<br />

each patient<br />

Before each<br />

fraction<br />

4. Patient-specific treatment<br />

verification (Image Guided)<br />

i. For SRS treatment<br />

ii. For SRT treatment<br />

1 mm<br />

2 mm<br />

Before<br />

treatment<br />

5. Safety Check<br />

Stereotactic interlocks Functional Daily<br />

6. Dosimetry<br />

Typical dose rate output constancy 2% Monthly<br />

Mechanical<br />

Treatment couch position indicators<br />

1 mm/0.5°<br />

7. Dosimetry<br />

i. SRS arc rotation mode<br />

(range: 0.5-10MU/deg)<br />

ii. X-ray monitor unit linearity<br />

(output constancy)<br />

Monitor units set vs. delivered:<br />

1 MU or 2% (whichever is greater)<br />

Gantry arc set vs. delivered:<br />

1° or 2% (whichever is greater)<br />

±5% for 2 - 4 MU<br />

±2% for ≥ 5 MU<br />

Annually<br />

Mechanical Checks<br />

i. Coincidence of radiation and<br />

mechanical isocentre<br />

ii. Stereotactic accessories, lockouts,<br />

etc.<br />

±1 mm from baseline<br />

Functional<br />

20


KEMENTERIAN KESIHATAN MALAYSIA<br />

Jadual 4.b: Additional QC Parameters for LINAC-Based<br />

Intensity Modulated Radiation Therapy (IMRT)<br />

No. Parameters Tolerance Frequency<br />

1. Individual patient plan verification ±5% (point dose), other per<br />

clinical significance<br />

Before first<br />

treatment<br />

2. Mechanical Check<br />

Laser localisation 1.5 mm Daily<br />

3. Dosimetry<br />

i. Typical dose rate output constancy 2% (@ IMRT dose rate) Monthly<br />

ii. Photon beam profile constancy 1%<br />

Mechanical Checks<br />

i. Distance check device for lasers<br />

compared with front pointer<br />

ii. Accessory trays (i.e., port film graticule<br />

tray)<br />

iii. Cross-hair centring (walkout)<br />

iv. Localising lasers<br />

1 mm<br />

2 mm<br />

1 mm (diameter)<br />

±1 mm<br />

*Safety Check<br />

Laser guard-interlock test<br />

Functional<br />

*Respiratory Gating<br />

Beam output constancy 2%<br />

i. Phase, amplitude beam control Functional<br />

ii. In room respiratory monitoring system<br />

Functional<br />

iii. Gating Interlock<br />

Functional<br />

4. Dosimetry<br />

i. X-ray flatness change from baseline 1% Annually<br />

ii. X-ray symmetry change from baseline ±1%<br />

iii. X-ray output calibration ±1% (absolute)<br />

iv.<br />

Spot check of field size dependent output<br />

factors for x ray(two or more FSs)<br />

v. X-ray beam quality (PDD 10 or<br />

vi.<br />

X-ray monitor unit linearity (output<br />

constancy)<br />

2% for field size


KEMENTERIAN KESIHATAN MALAYSIA<br />

No. Parameters Tolerance Frequency<br />

Mechanical Checks<br />

Table travel maximum range movement in all<br />

directions<br />

±2 mm<br />

Safety Checks<br />

Follow manufacturer’s test procedures<br />

Functional<br />

*Respiratory gating<br />

Beam energy constancy 2%<br />

Temporal accuracy of phase/amplitude gate<br />

on<br />

100ms of expected<br />

Calibration of surrogate for respiratory<br />

phase/amplitude<br />

100ms of expected<br />

Interlock testing<br />

Functional<br />

22


KEMENTERIAN KESIHATAN MALAYSIA<br />

Jadual 4.b.i: Additional QC Parameters for Multi-leaf Collimation<br />

(With differentiation of IMRT vs. non-IMRT machines)<br />

No. Parameters Tolerance Frequency<br />

1. Qualitative test (i.e., matched<br />

segments, also known as ―picket<br />

fence‖)<br />

Visual inspection for discernable<br />

deviations from expected leaf<br />

position<br />

Monthly<br />

(IMRT<br />

machines)<br />

2. i. Setting vs. radiation field for two<br />

patterns (non-IMRT)<br />

ii. Travel speed (IMRT for sliding<br />

window technique)<br />

iii. Leaf position accuracy (IMRT)<br />

2 mm Monthly<br />

Loss of leaf speed > 0.5cm/s<br />

1 mm for leaf positions of an<br />

IMRT field for four cardinal gantry<br />

angles*<br />

3. i. MLC transmission (average of leaf<br />

and interleaf transmission), all<br />

energies<br />

ii. Leaf position repeatability<br />

iii. MLC spoke shot<br />

iv. Coincidence of light field and x-ray<br />

field (all energies)<br />

v. Segmental IMRT (step and shoot)<br />

test<br />

vi. Moving window IMRT (four cardinal<br />

gantry angles)<br />

±0.5% from baseline Annually<br />

±1.0 mm<br />

≤1.0 mm radius<br />

±2.0 mm<br />


KEMENTERIAN KESIHATAN MALAYSIA<br />

Jadual 4.b.ii: Additional QC Parameters for Dynamic/universal/virtual wedges<br />

No. Parameters Tolerance Frequency<br />

1. Morning check-out run for<br />

one angle<br />

Dynamic Universal Virtual<br />

N/A* Functional N/A* Daily<br />

2. Wedge factor for all energies Central axis<br />

45º or 60º<br />

WF (within<br />

2%)<br />

Central axis<br />

45º or 60º<br />

WF (within<br />

2%)<br />

5% from<br />

unity,<br />

otherwise<br />

2%<br />

Monthly<br />

3. Check of wedge angle for<br />

60º, full field spot check for<br />

intermediate angle<br />

Check of off-centre ratios @ 80% field width<br />

@ 10 cm to be within 2%<br />

Annually<br />

*N/A – Not applicable<br />

24


KEMENTERIAN KESIHATAN MALAYSIA<br />

Jadual 5: Helical Tomotherapy<br />

No. Parameters Tolerance Frequency<br />

1. i. Output – rotational or static 3% Daily<br />

ii. Image/laser coordinate coincidence<br />

iii. Image registration/alignment<br />

iv. Red laser initialization<br />

2 mm for non-SRS/SBRT<br />

1 mm for SRS/SBRT<br />

1mm<br />

1.5 mm for non-SRS/SBRT<br />

1 mm for SRS/SBRT<br />

Radiation Safety Checks<br />

i. Door Interlock Functional<br />

ii. Audiovisual monitor<br />

Functional<br />

2. Beam parameters<br />

Output—Static 2% Monthly<br />

Output—Rotational 2%<br />

Monitor chamber constancy 2%<br />

Rotation output variation 2%<br />

Beam quality<br />

Transverse profile<br />

Longitudinal profiles (each slice width)<br />

1% PDD 10 or<br />

20<br />

TMR<br />

10<br />

1% average difference in field core<br />

1% of slice width at FWHM<br />

Alignment and Miscellaneous<br />

Interrupted procedure<br />

3%<br />

(agreement with uninterrupted<br />

procedure)<br />

Red laser movement<br />

1 mm<br />

Treatment couch<br />

i. Digital readout vs. actual movement 1 mm<br />

ii. Level 0.5°<br />

iii. Longitudinal motion alignment<br />

1 mm<br />

iv. Sag<br />

5 mm<br />

MVCT<br />

Geometric distortions<br />

Noise<br />

Uniformity<br />

Spatial resolution<br />

Contrast<br />

2 mm for non-SRS/SBRT<br />

1 mm for SRS/SBRT<br />

Consistency with baseline<br />

Consistency with baseline<br />

1.6 mm object<br />

Consistency with baseline<br />

25


KEMENTERIAN KESIHATAN MALAYSIA<br />

No. Parameters Tolerance Frequency<br />

(if MVCT is used for dose calculation)<br />

Uniformity<br />

HU (water test plug)<br />

HU (lung/bone test plug)<br />

25HU<br />

within ±HU 30 of baseline<br />

within ±HU 30 of baseline<br />

3. Machine Alignment<br />

i. MLC Tongue & Groove Out of focus percentage < 2% Annually<br />

ii. Jaw Shift<br />

Calculated IEC Y source<br />

position=±0.3 mm<br />

iii. Central axis Y-Axis misalignment<br />

Jaw offset =± 0.5 mm<br />

iv. MLC Centre-of-Rotation<br />

MLC COR Offset =±1.5mm<br />

MLC twist angle < 0.5 deg<br />

v. Field Centre vs. Jaw Setting Maximum field centre difference <<br />

0.5 mm<br />

vi. Jaw Detector Alignment<br />

Centre (IEC-Y) Offset =±2mm<br />

[left offset-right offset] < 2 mm<br />

4. Laser Alignment<br />

i. Green Overhead Laser Level X-axis divergence < 2 mm over 550<br />

mm vertical displacement<br />

Y-axis divergence < 2 mm over 550<br />

mm vertical displacement<br />

Annually<br />

ii. Green Overhead Laser Alignment<br />

Vertical offset = ±1 mm<br />

Vertical distance


KEMENTERIAN KESIHATAN MALAYSIA<br />

No. Parameters Tolerance Frequency<br />

v. Red Movable Lasers Red-green laser overlap ±1 mm<br />

Red-green laser twist


KEMENTERIAN KESIHATAN MALAYSIA<br />

No. Parameters Tolerance Frequency<br />

c. PDD curve For all field widths commissioned,<br />

PDD (IEC Z) profiles are within 2%<br />

of beam model values created at the<br />

factory from 10mm to 200mm.<br />

iii. Static Output & Energy<br />

Static output: ± 2% of calibrated<br />

output during commissioning<br />

Energy: ± 2% of factory data<br />

iv. Monitor Unit Calibration:<br />

(Treatment Beam and TomoImage<br />

beam)<br />

Measured vs. Displayed (Dose<br />

Monitor 1 & 2) Rates: within 2%<br />

Dose monitor 1 vs. Dose monitor 2:<br />

within 2%<br />

Expected vs. Cumulative MU: within<br />

2% of each other<br />

7. IMRT Verification<br />

Tomohelical Plan<br />

i. Plan 1:FW50mm;On-Axis Tumour<br />

ii. Plan 2:FW50mm;On-Axis Tumour<br />

iii. Plan 3:FW25mm;On-Axis Tumour<br />

iv. Plan 4:FW25mm;On-Axis Tumour<br />

v. Plan 5:FW10mm;On-Axis Tumour<br />

vi. Plan 6:FW10mm;On-Axis Tumour<br />

Measured doses are within 3% or 3<br />

mm of values predicted by beam<br />

model<br />

Annually<br />

Tomodirect Plan*<br />

i. Plan 1:FW50mm;On-Axis Tumour<br />

ii. Plan 2:FW50mm;On-Axis Tumour<br />

iii. Plan 3:FW25mm;On-Axis Tumour<br />

iv. Plan 4:FW25mm;On-Axis Tumour<br />

v. Plan 5:FW10mm;On-Axis Tumour<br />

vi. Plan 6:FW10mm;On-Axis Tumour<br />

Measured doses are within 4% or 3<br />

mm of values predicted by beam<br />

model<br />

8. TomoImage Verification<br />

i. Image Reconstruction Image is reconstructed and<br />

viewable on the Registration Panel<br />

Annually<br />

ii. Image Resolution<br />

The largest three rows of holes on<br />

the resolution plugs are visible.<br />

Each hole in the largest three rows<br />

can be distinguished from its<br />

neighbours<br />

28


KEMENTERIAN KESIHATAN MALAYSIA<br />

No. Parameters Tolerance Frequency<br />

iii. Image Artefacts<br />

iv. Dose<br />

There are no significant ring, streak,<br />

and button artefacts in the image<br />

Dose measured for a Fine<br />

TomoImage scan


KEMENTERIAN KESIHATAN MALAYSIA<br />

Jadual 6: Cyberknife<br />

No. Parameters Tolerance Frequency<br />

1. i. System status Passed/Failed Daily<br />

ii. Safety checks<br />

a) Door interlock (beam off)<br />

b) Audiovisual monitor<br />

iii. Collimator assembly collision detector<br />

iv. Detection of incorrect and missing<br />

collimator<br />

Functional<br />

Functional<br />

Functional<br />

Functional<br />

v. LINAC output constancy check < 2%<br />

vi. Robot mastering (perch position<br />

< 1mm<br />

laser) check<br />

vii. AQA targeting reproducibility test<br />

< 1mm from baseline<br />

2. i. Absolute dose 2% Monthly<br />

ii.<br />

X-ray energy<br />

a. Depth of d max<br />

b. Percent depth dose at D 10cm (PDD 10 )<br />

± 2mm from baseline (40mm<br />

collimator)<br />

± 2% from baseline<br />

(40mm collimator)<br />

iii. Flatness<br />

< 18% (40mm collimator)<br />

iv. Symmetry<br />

< 2% (40mm collimator)<br />

v. Penumbra < 4.5mm (40mm collimator)<br />

vi. LINAC/radiation field alignment<br />

< 1mm at 800mm SAD<br />

vii. Alignment of room lasers<br />

≤ 2mm<br />

viii. Imaging system alignment<br />

≤ 1mm of centre crosshairs<br />

or ± 2 pixels<br />

ix. Visual targeting test<br />

Laser on isocrystal for each<br />

node<br />

x. Treatment couch positioning check<br />

a. Head Up/Down at Home<br />

b. Roll Left/Right at Home<br />

c. Left/Right at Home<br />

0 ± 0.3 0<br />

0 ± 0.3 0<br />

< 5mm from centre<br />

30


KEMENTERIAN KESIHATAN MALAYSIA<br />

No. Parameters Tolerance Frequency<br />

xi. Imaging system bad pixel statistics Bad pixels less than<br />

maximum limit, number and<br />

position<br />

xii. TLS tracking and couch movement<br />

correspondence<br />

a. Left/Right ± 2mm<br />

b. Anterior/Posterior ± 2mm<br />

c. Inferior/Superior ± 2mm<br />

d. Roll Left/Right ± 0.3º<br />

e. Head Up/Down ± 0.3º<br />

f. RMS error ≤ 2mm<br />

Quarterly<br />

xiii. Treatment delivery targeting accuracy<br />

(End to End Test) (Total targeting error)<br />

< 0.95mm or < 1.5mm for<br />

Motion Tracking System<br />

xiv. CT geometric accuracy<br />

≤ 1mm<br />

xv. Treatment delivery dose accuracy < 5% (Typically about ± 2%)<br />

xvi. Iris collimator aperture size check<br />

Yes/No<br />

3. i. Commissioning beam data spot checks<br />

(Water Phantom measurements)<br />

a. X-ray energy See 2 (ii) above Annually<br />

b. Absolute dose 2%<br />

c. Dose rate As per LINAC specification<br />

d. Tissue Phantom Ratio (TPR) 2% from baseline<br />

e. Off Centre Ratio (OCR) 2% from baseline<br />

f. Output Factor (OF) 2% from baseline<br />

g. Flatness See 2 (iii) above<br />

h. Symmetry See 2 (iv) above<br />

i. Penumbra See 2 (v) above<br />

ii. Collimator transmission 1%<br />

iii. Dosimetry precision ± 2%<br />

iv. Linearity<br />

a. Requested MU = 10<br />

b. Requested MU = 20<br />

c. Requested MU = 30<br />

d. Requested MU = 40<br />

e. Requested MU = 50<br />

f. Requested MU = 100 to 1000<br />

± 6%<br />

± 4%<br />

± 3%<br />

± 2%<br />

± 1.5%<br />

± 1%<br />

31


KEMENTERIAN KESIHATAN MALAYSIA<br />

No. Parameters Tolerance Frequency<br />

v. Linac output rotational stability test ± 2%<br />

vi. TPS beam data spot checks<br />

± 2% of measured beam<br />

data<br />

vii. Interlocks and safety system tests<br />

Functional<br />

viii. Data security and verification<br />

Functional<br />

32


KEMENTERIAN KESIHATAN MALAYSIA<br />

Jadual 7: Brachytherapy<br />

Jadual 7.a: Remote Afterloading Brachytherapy Unit<br />

No. Parameters Tolerance Frequency<br />

1. i. Room safety door interlocks, lights, and<br />

alarms<br />

ii. Console functions, switches, batteries,<br />

printers<br />

iii. Visual inspection of source guides<br />

iv. System self test<br />

Functional<br />

Functional<br />

Free of kinks and<br />

firmly attached<br />

Functional<br />

Each treatment day<br />

2. i. Source positioning<br />

ii. Timer function<br />

1 mm<br />

1%<br />

Each treatment day<br />

or Weekly<br />

iii. Check accuracy of source guides and<br />

connectors<br />

1 mm<br />

3. Source strength calibration* 3% Each source change<br />

or Annually<br />

4. Accuracy of source and dummy loading<br />

(dummies used for spacing and/or simulation/<br />

verification)<br />

1 mm Annually<br />

5. Simulate emergency conditions Functional Annually<br />

6. Verify source inventory Performed Annually<br />

* It is worthwhile at source change to calibrate both new and old sources to establish and document<br />

reproducibility of calibration method.<br />

33


KEMENTERIAN KESIHATAN MALAYSIA<br />

Jadual 7.b: Special Procedures for Permanent Interstitial Brachytherapy<br />

No. Parameters Tolerance Frequency<br />

1. Safety and Radiation Protection<br />

i. Emergency equipment (forceps, tweezers,<br />

emergency safe, survey meter)<br />

Available<br />

ii. Source preparation area survey Performed<br />

Before the start of<br />

implant<br />

iii. Survey monitor Functional Before the start of<br />

implant or Annually<br />

iv. Hand held monitor<br />

Functional<br />

v. Protective material, lead aprons Functional<br />

vi. Source inventory Available Semi-Annually<br />

vii. Radiation survey for stray seeds Performed Upon completion of<br />

procedure<br />

viii. Monitor sources lost from the patient Performed Upon discharge<br />

2. Physical Parameters<br />

i. Source calibration, mean of batch<br />

ii. Source calibration, individual source<br />

iii. Source identification<br />

Maximum 3%<br />

Maximum 5%<br />

Verification<br />

Before the start of<br />

implant<br />

iv.<br />

Grid calibration in ultrasound system<br />

Performed<br />

Before the start of<br />

implant or<br />

Quarterly<br />

34


KEMENTERIAN KESIHATAN MALAYSIA<br />

Jadual 8: Treatment Planning System for Teletherapy<br />

No. Parameters Tolerance Frequency<br />

1. Checksum of <strong>program</strong> files no change Weekly<br />

2. Photon beam dosimetry<br />

i. Monitor unit calculation 1% Monthly<br />

ii. Standard treatment plan outputs 1% Quarterly<br />

iii. Point dose calculations and factor (PDD,TMR<br />

2% Annually<br />

etc for open and irregular fields)<br />

iv. Beam profiles (including open, wedged and<br />

2% Annually<br />

asymmetric beams)<br />

v. Inverse square law correction 2% Annually<br />

vi. Wedge and transmission factors 2% Annually<br />

3. Electron beam dosimetry<br />

i. Monitor unit calculation 1% Monthly<br />

ii. Standard treatment plan outputs 1% Quarterly<br />

iii. Point dose calculations and factors (PDD for<br />

2% Annually<br />

open and irregular fields)<br />

iv. Beam profiles 2% Annually<br />

v. Inverse square law correction 2% Annually<br />

vi. Beam size 2 mm Annually<br />

4. Peripheral devices<br />

i. Printing/Plotting device/ Film scanner/ Block<br />

cutting device/ Digitizer<br />

a. transfer of geometric figures Intact Monthly<br />

b. transfer of geometric dimensions 1 mm Monthly<br />

c. transfer of geometric orientations Intact Monthly<br />

ii. CT Interface<br />

a. transfer of geometric figures Intact Monthly<br />

b. transfer of geometric dimensions 1 mm Monthly<br />

c. transfer of geometric orientations Intact Monthly<br />

d. electron density compared to known densities<br />

in water<br />

10% Monthly<br />

iii. Archiving and retrieving of patient data Intact Semi-annually<br />

iv. Network data transfer Intact Semi-annually<br />

35


KEMENTERIAN KESIHATAN MALAYSIA<br />

Jadual 9: Imaging Systems<br />

(Portal Imaging and Image Guided Radiation Therapy)<br />

No. Parameters Tolerance<br />

Non-SRS/SBRT<br />

1. Planar kV and MV (EPID) imaging<br />

Tolerance<br />

SRS/SBRT<br />

Frequency<br />

i. Collision interlocks Functional Functional Daily<br />

ii. Positioning/repositioning ≤2 mm ≤1 mm Daily<br />

iii. Imaging and treatment coordinate<br />

coincidence (single gantry angle)<br />

≤2 mm ≤1 mm Weekly<br />

Cone-beam CT (kV and MV)<br />

i. Collision interlocks Functional Functional Daily<br />

ii. Positioning/repositioning ≤1 mm ≤1 mm Daily<br />

iii. Imaging and treatment coordinate<br />

coincidence<br />

≤2 mm ≤1 mm Weekly<br />

2. Planar MV imaging (EPID)<br />

i. Imaging and treatment coordinate<br />

coincidence (four cardinal angles)<br />

≤2 mm ≤1 mm Semi-<br />

Annually<br />

ii. Scaling ≤2 mm ≤2 mm<br />

iii. Spatial resolution Baseline* Baseline<br />

iv. Contrast Baseline Baseline<br />

v. Uniformity and noise Baseline Baseline<br />

Planar kV imaging<br />

i. Imaging and treatment coordinate ≤2 mm<br />

≤1 mm<br />

coincidence (four cardinal angles)<br />

ii. Scaling ≤2 mm ≤1 mm<br />

iii. Spatial resolution Baseline Baseline<br />

iv. Contrast Baseline Baseline<br />

v. Uniformity and noise Baseline Baseline<br />

Cone-beam CT (kV and MV)<br />

i. Geometric distortion ≤2 mm ≤1 mm<br />

ii. Spatial resolution Baseline Baseline<br />

iii. Contrast Baseline Baseline<br />

iv. HU constancy Baseline Baseline<br />

v. Uniformity and noise Baseline Baseline<br />

36


KEMENTERIAN KESIHATAN MALAYSIA<br />

No. Parameters Tolerance<br />

Non-SRS/SBRT<br />

Tolerance<br />

SRS/SBRT<br />

Frequency<br />

3. Planar MV imaging (EPID)<br />

Full range of travel SDD ±5 mm ±5 mm Annually<br />

Imaging dose Baseline Baseline<br />

Planar kV imaging<br />

Beam quality/energy Baseline Baseline<br />

Imaging dose Baseline Baseline<br />

Cone-beam CT (kV and MV)<br />

Imaging dose Baseline Baseline<br />

*Baseline means that the measured data are consistent with acceptance testing data.<br />

37


KEMENTERIAN KESIHATAN MALAYSIA<br />

Jadual 10: Intra-operative Radiation Therapy (Low kV Photon)<br />

No. Parameters Tolerance Frequency<br />

1. Radiation Safety<br />

i. Radiation on/off warning light<br />

ii. Survey meter<br />

iii. Mobile Lead Shield<br />

iv. X-ray source calibration certificate<br />

v. Ionization chamber calibration certificate<br />

Functional<br />

Available and Functional<br />

Available<br />

Valid<br />

Valid<br />

Before each<br />

treatment<br />

day<br />

2. Infection Safety<br />

Sterilization * for Interstitial Treatment:<br />

i. Photodiode array<br />

ii. X-ray source<br />

iii. Probe adjuster/ion chamber holder<br />

iv. External Radiation Monitor (if used)<br />

v. X-ray source and probe adjuster/ion<br />

chamber holder cables<br />

vi. Verification block (V block)<br />

vii. X-ray source holding block (X block)<br />

viii. Sterilization tray<br />

Manufacturer’s<br />

specifications<br />

Before each<br />

case<br />

3. Quality Assurance System<br />

i. Integrity of applicators (visual check)<br />

ii. X-ray probe straightness<br />

iii. Electron beam dynamic offset<br />

iv. Source isotropy<br />

v. Internal Radiation Monitor<br />

vi. Source dose rate (Gy/min)<br />

Satisfactory<br />

< 0.1 mm<br />

< 1 mm<br />

< 10 %<br />

+5 % (from initial value)<br />

< 5 % from certificate<br />

Before each<br />

case or<br />

Monthly<br />

4. Radiation Safety Checks<br />

i. Ion chamber calibration<br />

ii. X-Ray Source Output Factor Calibration<br />

iii. Percent Depth Dose for all applicators<br />

iv. Radiation anisotropy at 10 mm from<br />

surface for all applicators<br />

v. Radiation leakage at 10 cm from properly<br />

shielded** source<br />

Local authority<br />

Manufacturer specification<br />

Manufacturer specification<br />

< 5 %<br />

< 0.5 %<br />

Annually<br />

* Sterilization not mandatory for tumour bed treatment<br />

** With PDA (photodiode array) in place<br />

38

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!