OPERATING PRINCIPLE<br />

The P-700’s measurement cycle is based on the ASTM Methods D-323, D-4953<br />

and D-5482 and correlates to D-5191 and D-6377. This is done by using a digitally<br />

controlled syringe sample handling system, micro-volume solenoid valves and an<br />

angled measurement chamber equipped with a high-resolution pressure sensor and<br />

magnetic stirrer.


73.0<br />

(1855)<br />

To Air<br />

Coolers<br />

Explosion Proof Unit<br />

37.0<br />

(940)<br />

Ø .50 (13) Hole<br />

4 Places<br />

37.00<br />

(940)<br />

Customer<br />

Connection<br />

#1<br />

1/2" FNPT<br />

(M20)<br />

Nitrogen Purged Unit<br />

22.00<br />

(559)<br />

24.00<br />

(610)


PRODUCT GUIDE<br />

Petroleum Analyzers<br />

•Flash Point<br />

•Salt In Crude<br />

•RVP<br />

•RVP/VL20<br />

•Freeze Point<br />

•Cloud Point<br />

•Pour Point<br />

•Viscosity<br />

•Viscosity Index<br />

Water Analyzers<br />

•UV-COD<br />

•UV-Oil in Water<br />

Other Products<br />

•Environmental Cabinets<br />

•Sample Conditioning<br />

Systems<br />

•Sample Recovery<br />

Systems<br />

•Spare Parts<br />

Analyzer Services<br />

•Field Service<br />

•Start-Ups<br />

•Training<br />

•Technical Support<br />

SPECIFICATIONS: P-700 RVP ANALYZER<br />

ANALYSIS PERFORMANCE<br />

Measurement Cycle Time 5 minutes or less<br />

Measurement Range 0 – 35 psi / 0 – 2.4 bar / 0 – 2400 mbar / 0 – 240 kPa / 0 – 2400 hPa (selectable)<br />

Repeatability ± 0.05 psi (± 0.0035 bar)<br />

Reproducibility ± 0.1 psi (± 0.007 bar)<br />

Resolution ± 0.01 psi (± 0.0007 bar)<br />

Accuracy Meets or exceeds ASTM Methods D-323, D-4953, D-5482, D-5191 & D-6377<br />

Pressure Accuracy ± 0.01% of full scale<br />

Temperature Accuracy<br />

SAMPLE REQUIREMENTS<br />

± 0.2°F (± 0.1°C) of full scale<br />

Sample Bypass Flow Rate Min. 0.04 L/min – Max. 0.1 L/min<br />

Sample Return Pressure<br />

Atmospheric – Max. 35 psi (2.4 bar) - optional high pressure sample recovery system<br />

available (P/N 700172)<br />

Sample Pressure<br />

Min. 20 psi (1.4 bar) – Max. 35 psi (2.4 bar) - optional sample conditioning system<br />

available (P/N 700173)<br />

Sample Temperature Min. 35°F (2°C) – Max. 105°F (40°C)<br />

Sample Particulates less than 10 µm - optional sample conditioning system available (P/N 700173)<br />

Sample Conditions homogenous, single-phase sample without water or water moisture<br />

ENCLOSURE/INSTALLATION REQUIREMENTS<br />

Dimensions (Purged) Width 24.0 in (610mm) – Height 39.0 in (991mm) – Depth 9.51 in (242mm)<br />

Dimensions (Exd) Width 37.0 in (940mm) – Height 73.7 in (1874mm) – Depth 30.0 in (762mm)<br />

Weight Purged Unit 150 lbs (68 kg)/ Exd Unit 500 lbs (228 kg)<br />

Operating Temperature Min. 40°F (5°C) – Max. 105°F (40°C)<br />

Enclosure Material/Rating stainless steel - NEMA 4X / IP65<br />

Area Classification NEC Class 1 Div 1 Group B, C + D or ATEX Zone1 II B + H2 T4<br />

Power self-selecting 100 to 120 VAC or 200 to 240 VAC, 50/60 Hz, single phase, 2A<br />

Cabinet Purge Gas Supply<br />

Clean, dry Nitrogen or other inert gas (better than 98% pure) at Min. 40 psi (2.7 bar)<br />

– Max. 100 psi (6.8 bar expected leakage <strong>com</strong>pensation 1 l/min<br />

Purge System Air Logic Supply Instrument grade air at Min. 40 psi (2.7 bar) – Max. 100 psi (6.8 bar)<br />

Cell Purge Gas Supply<br />

END USER CONNECTIONS<br />

Clean, dry Nitrogen or other inert gas (better than 98% pure) at Min. 20 psi (1.4 bar)<br />

– Max. 35 psi (2.4 bar) / approx. 20 ccm/min flow rate at 20% duty cycle<br />

Analog Output Signal<br />

single isolated 4-20 mA output (optional second output available), selectable for sample<br />

RVP values, analyzer system/maintenance warning or analysis measurement indication<br />

Relay Output Contact<br />

three SPDT Relays with contacts rated at 3A resistive load at 250VAC ,selectable for<br />

sample RVP value alarm, analyzer maintenance warning or analyzer fault alarm<br />

Serial Input/Output Signal single RS232 or RS485 bi-directional / optional ModBus output available<br />

HOW TO ORDER<br />

ANALYZER SYSTEMS<br />

Catalog Number P-700-1100 <strong>ORB</strong> Model P-700 RVP Analyzer, Ex Area ready for NEC Class 1 Div 1 Group C + D<br />

Catalog Number P-700-1200 <strong>ORB</strong> Model P-700 RVP Analyzer, Ex Area ready for ATEX Zone1 II B + H2 T4<br />

Catalog Number P-700-1400 <strong>ORB</strong> Model P-700 RVP Analyzer, NEC Class 1 Div 1 Group B, C + D<br />

OPTIONS<br />

Catalog Number 700170 Validation/Grab Sample System, Micro Flow<br />

Catalog Number 700171 Dual-Stream Sampling System, Micro Flow<br />

ACCESSORIES<br />

Catalog Number 700174-P700 Free-standing Mounting Rack<br />

Catalog Number 700175 1-Year Spare Parts Kit<br />

Catalog Number 700176 2-Year Spare Parts Kit<br />

<strong>ORB</strong> <strong>Instruments</strong>, <strong>Inc</strong>.<br />

4724 South Christiana<br />

Chicago, IL 60632 / USA<br />

Phone: + (1) 773 927-8600<br />

Fax: + (1) 773 927-8620<br />

Email: sales@orbinstruments.<strong>com</strong><br />

www.orbinstruments.<strong>com</strong><br />

Lit. No. P-700-EN-US / JAN10