8 TROUBLESHOOTING<br />

WARNING: If you experience a problem with your <strong>lift</strong> and are unable to service it yourself,<br />

contact your Lumex dealer.<br />

LF1030 HYDRAULIC PATIENT LIFT<br />

SYMPTOM POSSIBLE CAUSE ACTION<br />

Patient <strong>lift</strong> does not go up<br />

or down<br />

Air in the cylinder<br />

Bleed the cylinder by turning the release<br />

knob and push down on the mast. Pump<br />

up the <strong>lift</strong> again and turn the release knob<br />

again.<br />

Loss of <strong>hydraulic</strong> fluid (leaking Replace any leaking cylinder<br />

cylinder)<br />

Patient <strong>lift</strong> does not go Air in the cylinder<br />

See above Directions<br />

down<br />

Note: If the <strong>hydraulic</strong> cylinder begins leaking fluid, do not operate. Contact your Lumex<br />

dealer for a new cylinder.<br />

LF1040 ELECTRIC PATIENT LIFT<br />

SYMPTOM POSSIBLE CAUSE ACTION<br />

Control box is not<br />

responding and relay is not<br />

clicking<br />

Actuator does not run, but<br />

relay is clicking<br />

Actuator goes up or down,<br />

but not both<br />

Battery discharged<br />

Battery charger defective<br />

Battery defective<br />

Hand control defective<br />

Actuator plug not properly seated<br />

Actuator defective<br />

Control box defective<br />

Hand control defective<br />

Control box defective<br />

Charge battery<br />

Replace battery charger<br />

Replace battery<br />

Replace hand control<br />

Push actuator plug firmly into control box<br />

Replace actuator<br />

Replace control box<br />

Replace hand control<br />

Replace control box<br />

9 SPECIFICATIONS *<br />

Note: All dimensions are given in inches unless otherwise specified.<br />

LF1030 HYDRAULIC PATIENT LIFT / LF1040 ELECTRIC PATIENT LIFT<br />

Base width Open: 39.4"<br />

Closed: 25.6"<br />

Base height Clearance 5-1/2"<br />

Base length 45.30"<br />

Lifting height range Note: As measured from point<br />

where spreader bar attaches to boom.<br />

Minimum: 27.4"<br />

Maximum: 77.7"<br />

Range: 50.3"<br />

Mast height 55.1"<br />

Weight capacity<br />

400 lb (181 kg)<br />

Caster size<br />

4.00" diameter, both locking and non-locking<br />

Product Weight<br />

90 lb<br />

Shipping Weight<br />

103.2 lb<br />

* Specifications are subject to change without notice.<br />