Bomem-GRAMS User's Guide
Bomem-GRAMS User's Guide
Bomem-GRAMS User's Guide
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Appendix B<br />
Array Basic Interface<br />
Array Basic methods<br />
OBJXCALL "TaskObj",<br />
"CONFIG_HARDWARE", , cardid, timeout<br />
OBJXCALL "TaskObj",<br />
"CONFIG_HARDWARE_ID", , bname, timeout<br />
OBJXCALL "TaskObj", "CONFIG_PROGRAM",<br />
, buffer, warn, format, save_option<br />
OBJXCALL "TaskObj",<br />
"CONFIG_SATURATION", , max_aplitude,<br />
sigma_min, sigma_max, threshold<br />
Sets hardware parameters<br />
Parameters<br />
cardid:<br />
Description<br />
The zero based index of the instrument to use for<br />
acquisition. The order may vary depending on the<br />
available instruments. It is recommended to use the<br />
"INSTRUMENT_INDEX" and "INSTRUMENT_ID"<br />
properties to determine the appropriate cardid value to<br />
use.<br />
timeout: the longest time in seconds that the driver will wait for a<br />
response before declaring an error<br />
Sets hardware parameters<br />
Parameters<br />
bname:<br />
the name (or description or ID) of the instrument to use<br />
for acquisition (as returned by the "INSTRUMENT_ID"<br />
property)<br />
timeout: the longest time in seconds that the driver will wait for a<br />
response before declaring an error<br />
Sets program parameters<br />
Parameters<br />
buffer:<br />
warn:<br />
format:<br />
this is the size of the FIFO buffer in kilobytes for holding<br />
spectra in memory. This parameters is normally set<br />
to 4096<br />
true of false field for the file overwrite warning. If true,<br />
the user must confirm file overwrites before they occur<br />
0 - Float IEEE (recommended)<br />
1 - New SPC <strong>GRAMS</strong> format<br />
save_option: 0 - to save full spectral range,<br />
1 - to save specified spectral range<br />
Sets saturation parameters<br />
Parameters<br />
max_amplitude:<br />
maximum allowed interferogram ZPD<br />
amplitude. This value is used for detecting<br />
saturation and computing percent of maximum<br />
in align mode<br />
sigma_min: start of out of band region in cm -1 for linearity and<br />
optical saturation test<br />
sigma_max: end of out if band region in cm -1 for linearity and<br />
optical saturation test<br />
threshold: highest allowed value for the ratio of the average band<br />
intensity with the maximum spectral intensity in<br />
percentage<br />
36 <strong>Bomem</strong>-<strong>GRAMS</strong> <strong>User's</strong> <strong>Guide</strong>