Appendix B<br />
Array Basic Interface<br />
Array Basic methods<br />
OBJXCALL "TaskObj", "START_SCAN",,<br />
sigma_min, sigma_max, delay, scans, name,<br />
memo, type, bname, use_autogain, use_display<br />
OBJXCALL "TaskObj", "START_KSCAN",,<br />
sigma_min, sigma_max, delay, scans, name,<br />
memo, type, bname, subfiles, sdelay,<br />
use_autogain, use_display<br />
Description<br />
Start a sample scan<br />
Parameters<br />
sigma_min: start of display range in cm -1<br />
sigma_max: end of display range in cm -1<br />
delay: delay in seconds before acquisition starts<br />
scans: number of scans to acquire<br />
name: file name to save data under, extension is always imposed<br />
according to the type of spectrum acquired<br />
memo: comment to associate with this file<br />
type: 0 - Interferogram, extension .interferogram.spc<br />
1 - Raw spectrum, extension .SingleBeam.spc<br />
2 - Transmittance, extension .Transmittance.spc<br />
3 - Absorbance, extension .Absorbance.spc<br />
bname: The file name of the background. This is used if the<br />
type is transmittance or absorbance<br />
use_autogain: true or false field for use of autogain<br />
use_display: true or false field for use of display during<br />
acquisition<br />
Start a sample scan in kinetic mode<br />
Parameters<br />
sigma_min: start of display range in cm -1<br />
sigma_max: end of display range in cm -1<br />
delay: delay in seconds before acquisition starts<br />
scans: number of scans to acquire<br />
name: file name to save data under, extension is always imposed<br />
according to the type of spectrum acquired<br />
memo: comment to associate with this file<br />
type: 0 - Interferogram, extension .interferogram.spc<br />
1 - Raw spectrum, extension .SingleBeam.spc<br />
2 - Transmittance, extension .Transmittance.spc<br />
3 - Absorbance, extension .Absorbance.spc<br />
bname: the file name of the background. This is used if the<br />
type is transmittance or absorbance<br />
subfiles: the number of consecutive measurements to acquire in<br />
this kinetic run<br />
sdelay: the delay between subfiles measured from start to start<br />
use_autogain: true or false field for use of autogain<br />
use_display: true or false field for use of display during<br />
acquisition<br />
<strong>Bomem</strong>-<strong>GRAMS</strong> <strong>User's</strong> <strong>Guide</strong> 35