Section 4.1<br />
Description<br />
b The recommended setting<br />
for this field is Cosine.<br />
• Apodization b selects the apodization function that is applied to<br />
the interferogram before computing the Fast Fourier Transform (FFT).<br />
The available apodization functions are listed and described in Appendix A<br />
on page 31.<br />
• Laser Freq must be set to the frequency of the reference laser.<br />
This value should be 15799.7 for the red He-Ne laser used in all standard FT-<br />
IR Series spectrometers. Changing this value will render the data acquired<br />
incompatible with other data and with standard methods and libraries.<br />
• Absorbance Cut-Off min and Cut-Off max define the minimum and<br />
maximum values that any absorbance spectrum can reach.<br />
If the spectrum goes out of this range it is automatically truncated.<br />
• Transmittance Cut-Off min and Cut-Off max define the minimum and<br />
maximum values that any transmittance spectrum can reach.<br />
If the spectrum goes out of this range it is automatically truncated.<br />
18 <strong>Bomem</strong>-<strong>GRAMS</strong> <strong>User's</strong> <strong>Guide</strong>