Section 1.3<br />

Acquire a sample spectrum<br />

1.3 Acquire<br />

a sample<br />

spectrum<br />

1. When the background spectrum has been acquired, insert the sample in<br />

the sample holder or sampling accessory.<br />

For an FT-IR series spectrometer:<br />

• For MIR, use the polystyrene sample supplied with<br />

the spectrometer (see Figure 1-8 below).<br />

• For NIR, do not use the polystyrene sample. Instead, place an<br />

ordinary plastic bag over the sample holder.<br />

For another model of spectrometer, use a sample of the appropriate type.<br />

Figure 1-8. Inserting the polystyrene sample into the sample holder of an MB Series spectrometer<br />

2. Select Collect to display the Collect dialog box (see Figure 1-7 on page 5).<br />

Set the parameters as follows:<br />

Acquisition type: Normal<br />

Spectral Range Start and End:<br />

• for MIR: from 4000 to 400 cm -1<br />

• for NIR: from 14 000 to 4000 cm -1<br />

General Param:<br />

• Initial Delay: 0<br />

• Number of scans: 10 or more<br />