Messages<br />
Messages which occur can be split into Warnings and Alarm messages.<br />
Please note that alarms generally cause the machine to stop temperature control. An<br />
alarm message is immediately displayed as text on the graphic display (60).<br />
After the reason for the alarm has been repaired, the machine must be switched off<br />
using the main switch (36), and then switched on again in order to reset the alarm.<br />
Warnings do not normally cause the machine to stop temperature control, however they<br />
give information concerning critical conditions within the machine or environment. If the<br />
conditions leading to the warning message are not improved, then there is a danger that<br />
the machine will stop with an alarm, as generally alarm limits are then exceeded.<br />
After the alarm or warning message has been acknowledged, then the graphic display<br />
(60) shows in the left upper corner a symbol (a stop sign for alarm messages, a triangle<br />
with an exclamation mark for warnings). By turning the rotary knob / key (61),<br />
alternatively by lightly touching the screen (with Unistat Pilot) the symbol can be<br />
selected. By pressing the rotary knob / key (61) a further information window opens up,<br />
in which the alarm or warning messages are chronologically shown. Turning the rotary<br />
knob / key (61) allows specific messages to be chosen.<br />