Mutual functions Com.G@te/Web.G@te<br />

PoKo (potential free contact) Alarm plug connector<br />

Signal contact for external monitoring<br />

Observe the functional options, which the PoKo provides in the main menu.<br />

The potential free contact (PoKo) signals the condition of the temperature control<br />

device by means of the contact position. A closed operating contact means ready<br />

status (must be set via the menu). The operating contact is open in case of a fault or<br />

error (this applies to the normally open contact between pin 1 and pin 2). Further<br />

applications are possible e.g checking of the actual value. The connection is designed<br />

as a potential free changeover contact.<br />

72<br />

1 3<br />

2<br />

ECS Socket (External Control Signal) Standby<br />

Release signal ECS (External Control Signal), for starting / stopping temperature control.<br />

The ECS can be used to switch of the temperature control mode (internal/process),<br />

switch to the 2 nd set-point or allow temperature control.<br />

Control over a potential free contact. The contacts 1 and 3 are bridged internally. ECS<br />

becomes electronically active, if E1 and E2 are connected by an external zero potential<br />

free contact.<br />

The functionality of the ECS is determined in the Com.G@te / Web.G@te menu under<br />

the point ECS Standby.<br />

The following options are offered:<br />

- OFF: ECS does not have any effect (Factory default)<br />

- 2 nd set-point: The 2 nd set-point is taken over as soon as ECS is activated<br />

- Standby: ECS has effect on temperature control (ON / OFF)<br />

3<br />

2<br />

1<br />

Normally open contact between pin 1 and pin 2.<br />

Normally closed contact between pin 2 and pin 3.<br />

Contact load: 1A at 24V DC<br />

Pin Signal<br />

1,3 E2<br />

2 E1