Mutual functions Com.G@te/Web.G@te<br />
PoKo (potential free contact) Alarm plug connector<br />
Signal contact for external monitoring<br />
Observe the functional options, which the PoKo provides in the main menu.<br />
The potential free contact (PoKo) signals the condition of the temperature control<br />
device by means of the contact position. A closed operating contact means ready<br />
status (must be set via the menu). The operating contact is open in case of a fault or<br />
error (this applies to the normally open contact between pin 1 and pin 2). Further<br />
applications are possible e.g checking of the actual value. The connection is designed<br />
as a potential free changeover contact.<br />
72<br />
1 3<br />
2<br />
ECS Socket (External Control Signal) Standby<br />
Release signal ECS (External Control Signal), for starting / stopping temperature control.<br />
The ECS can be used to switch of the temperature control mode (internal/process),<br />
switch to the 2 nd set-point or allow temperature control.<br />
Control over a potential free contact. The contacts 1 and 3 are bridged internally. ECS<br />
becomes electronically active, if E1 and E2 are connected by an external zero potential<br />
free contact.<br />
The functionality of the ECS is determined in the Com.G@te / Web.G@te menu under<br />
the point ECS Standby.<br />
The following options are offered:<br />
- OFF: ECS does not have any effect (Factory default)<br />
- 2 nd set-point: The 2 nd set-point is taken over as soon as ECS is activated<br />
- Standby: ECS has effect on temperature control (ON / OFF)<br />
3<br />
2<br />
1<br />
Normally open contact between pin 1 and pin 2.<br />
Normally closed contact between pin 2 and pin 3.<br />
Contact load: 1A at 24V DC<br />
Pin Signal<br />
1,3 E2<br />
2 E1