70 � LAI Command Set The master queries whether the slave 01 is connected to the bus. Master query: [M01V07C6\r Command structure: [ Start signal (5Bh) 1 Byte M transmitter identification M (4Dh) for master or S (53h) for slave 1 Byte 01 Slave address 01…99 2 Byte V Identifier data group 1 Byte 07 Length of the data field 2 Byte C6 Check sum 2 Byte \r End-character CR (0Dh) 1 Byte In order to increase the security of the data, a check sum is transmitted. The check sum is the 1 byte sum of all hex values from the start character to the last character in front of the check sum. Slave answer: [S01V0EMINI CCAD\r Slave 01 is connected to the bus, and the device is a MINI CC. The slave command has the „MINI CC“ data group, which is 7 bytes long. These 7 bytes plus the 7 bytes in front of the data group produce a data field length of 14 bytes = 0Eh byte. � Namur-Commands IN_PV_00\r\n Request temperature Tinternal (jacket) IN_PV_02\r\n Request temperature Texternal (proess) IN_SP_00\r\n Request temperature set-point IN_SP_05\r\n Request current analogue set-point OUT_SP_00\r\n Send set-point xxx.xx START\r\n Start temperature control STOP\r\n Stop temperature control STATUS\r\n Request device status STATUS Single digit -1 Alarm / Warning Manual Stop 0 OK / standby / manual stop Manual Stop 1 OK Temperature control / air-purging Remote Stop 2 Temperature control stopped remotely Remote Stop 3 Temperature control started remotely Data flow control: The commands are not buffered. A new command may be sent as soon as the reply to the previous command has been received. If no reply is expected, then there should be a pause of 500 ms.

The following interfaces are available: 1. RS232/SERIAL (Standard Interface! Included in all CC units with CC-Pilot) 2. Com.G@te internal or external version depending on the unit (Extension of an analogue interface or a further digital interface) 3. Web.G@te internal or external version depending on the unit (Control possible via Intranet, Internet, USB…) Caution! With plug-in connections subject to a higher voltage than 60 V (AC) and 40 V (DC) there is the risk of receiving an electric shock or of damaging the interface module. Only components requiring voltages of less than 60 V (AC) and 40 V (DC) should be connected, in order to ensure the safety of the interface module. RS232/SERIAL Wiring RS232: Pin2 RxD Receive Data Pin3 TxD Transmit Data Pin5 GND Signal GND The Baudrate is set to 9600. It cannot be changed! An operation via the interface RS232/SERIAL is not possible if the temperature control unit is controlled via remote control by the CC-Pilot. 5 9 6 1 71

70<br />

� LAI Command Set<br />

The master queries whether the slave 01 is connected to the bus.<br />

Master query: [M01V07C6\r<br />

Command structure:<br />

[ Start signal (5Bh) 1 Byte<br />

M transmitter identification M (4Dh) for master or<br />

S (53h) for slave 1 Byte<br />

01 Slave address 01…99 2 Byte<br />

V Identifier data group 1 Byte<br />

07 Length of the data field 2 Byte<br />

C6 Check sum 2 Byte<br />

\r End-character CR (0Dh) 1 Byte<br />

In order to increase the security of the data, a check sum is transmitted. The check sum<br />

is the 1 byte sum of all hex values from the start character to the last character in front<br />

of the check sum.<br />

Slave answer: [S01V0EMINI <strong>CC</strong>AD\r<br />

Slave 01 is connected to the bus, and the device is a MINI <strong>CC</strong>. The slave command has<br />

the „MINI <strong>CC</strong>“ data group, which is 7 bytes long. These 7 bytes plus the 7 bytes in<br />

front of the data group produce a data field length of 14 bytes = 0Eh byte.<br />

� Namur-Commands<br />

IN_PV_00\r\n Request temperature Tinternal (jacket)<br />

IN_PV_02\r\n Request temperature Texternal (proess)<br />

IN_SP_00\r\n Request temperature set-point<br />

IN_SP_05\r\n Request current analogue set-point<br />

OUT_SP_00\r\n Send set-point xxx.xx<br />

START\r\n Start temperature control<br />

STOP\r\n Stop temperature control<br />

STATUS\r\n Request device status<br />

STATUS Single digit<br />

-1 Alarm / Warning<br />

Manual Stop 0 OK / standby / <strong>manual</strong> stop<br />

Manual Stop 1 OK Temperature control / air-purging<br />

Remote Stop 2 Temperature control stopped remotely<br />

Remote Stop 3 Temperature control started remotely<br />

Data flow control: The commands are not buffered. A new command may be sent as<br />

soon as the reply to the previous command has been received. If no reply is expected,<br />

then there should be a pause of 500 ms.