Setting the set-point limits<br />
The minimum and maximum set-point limits provide safety for the equipment. These<br />
limits should be set before starting temperature control and when changing the thermal<br />
fluid in relation to the temperature range of the thermal fluid.<br />
The maximum set-point limits the set-point input of the outlet temperature. The<br />
minimum set-point protects against viscosity or freezing of the thermal fluid with low<br />
temperatures. The set-point input is only possible in the band between maximum and<br />
minimum set-point limit.<br />
Select the Set-point limits option from the Main menu. Enter the required minimum /<br />
maximum set-point using the rotary knob /key (61), and then confirm it by pressing the<br />
rotary knob / key (61).<br />
Check the value of the minimum and maximum set-point for every system change, in<br />
particular if the thermal fluid is changed.<br />
Entering a set-point<br />
Select the Set-point option from the Main menu. The new set-point can be chosen, and<br />
confirmed using the rotary knob (61). The value of the set-point is limited by the current<br />
minimum and maximum set-point limits.<br />
The following is true:<br />
minimum set-point