Setting the safety value:<br />

Select the Over-temperature protection from the main menu. Choose the menu point<br />

OT limits: heating. A code generator (random generator) will give out a code via the<br />

graphics display. This code has to be entered when requested (a deliberate and knowing<br />

action). This procedure replaces the usual setting by means of a tool (screw driver).<br />

Afterwards the new safety value can be entered.<br />

Setting the OT Mode:<br />

Information! Standard setting is Stop to DIN EN 61010. Please note that<br />

after a default setting the over-temperature protection will be set to Stop<br />

according to DIN EN 61010!<br />

Select the Over-temperature protection from the main menu. Choose the menu point<br />

OT mode. A code generator (random generator) will give out a code via the graphics<br />

display. This code has to be entered when requested (a deliberate and knowing action).<br />

This procedure replaces the usual setting by means of a tool (screw driver). Afterwards<br />

the new safety value can be entered.<br />

Stop according to DIN EN 61010<br />

When the cut-off temperature is reached, heating, cooling and pump are all turned off<br />

(Default setting).<br />

Cut-off mode: Process Safety<br />

When the cut-off temperature is reached the heating is turned off, but the cooling (at<br />

maximum power) and pump remain on. The full cooling capacity of the unit remains<br />

available for emergency cooling (possible exothermic reaction). Please ensure that the<br />

compressor automatic is switched to always on using the menu point Compressor<br />

automatic (F35). Please also see description on chapter Comfort menu for the<br />

respective Function numbers and their meanings.<br />

Display all OT values:<br />

The present measured temperature of the over-temperature sensors, the trip<br />

temperature set and the cut-off mode are displayed.<br />

Please note, that some units use two over-temperature sensors. When selecting Display<br />

all OT values both values will be displayed.<br />