Connecting an externally closed application (reactor)<br />
Remove the thread covers from the unit´s circulation flow (1) and circulation return (2).<br />
Make sure that the hose material is compatible with the thermal fluid and temperature<br />
range being used. In order that the application can be driven correctly, and that no air<br />
bubbles remain in the system, ensure that the unit´s circulation flow (1) is attached to<br />
the lowest connection on the application, and the unit´s circulation return (2) is<br />
attached to the highest connection on the application.<br />
Please note the markings for the hose connections on the housing<br />
60<br />
2<br />
1<br />
Markings on the<br />
temperature control<br />
device