COM.G@TE allocation and setting (quick guide)<br />

6<br />

1<br />

2<br />

3<br />

5<br />

4<br />

6<br />

3<br />

2<br />

1<br />

3<br />

2<br />

1<br />

5<br />

1<br />

9 6<br />

POCO (potential free contact) Alarm plug-in connector<br />

Signal contact for external monitoring.<br />

The connection is designed as a potential free changeover contact.<br />

Normally open contact between pin 1 and pin 2. Normally closed contact between pin 2 and pin 3.<br />

Contact load: 1A at 24V DC Only use screened lines!<br />

AIF Reg-E-Prog Socket<br />

Analogue interface, one input channel (programmable) and 3 output channels.<br />

Pin Signal<br />

1. Current output, T extern 0/4-20mA or 0-10V<br />

2. Current output, set-point 0/4-20mA or 0-10V<br />

3. GND for analogue outputs GND<br />

4. Analogue input (programmable) 0/4-20mA or 0-10V<br />

5. Current output, free programmable 0/4-20mA or 0-10V<br />

6. GND for analogue input GND<br />

ECS Socket (External Control Signal) Standby<br />

Release signal ECS (External Control Signal), for starting / stopping temperature control.<br />

The following variants are offered:<br />

Pin Signal<br />

1,3 E2<br />

2 E1<br />

RS232 / RS485 Serial Socket<br />

Wiring RS232: Wiring RS485<br />

Pin2 RxD Receive Data Pin6 A with 120 Ohm load resistance<br />

Pin3 TxD Transmit Data Pin7 A<br />

Pin5 GND Signal GND Pin8 B<br />

Functions in connection with PLS<br />

Settings for analogue interface, release signal, signal contact and digital interface RS232 / RS485 can be made via the<br />

functions Analogue Interface, RS232 / RS485, ECS Standby and POCO Alarm in the COM.G@TE menu.<br />

Standard settings are:<br />

Analogue interface: analogue input OFF and analogue output OFF<br />

RS232 / RS485: RS232 with Baudrate 9600<br />

ECS Standby: no action<br />

POCO Alarm: no alarm<br />

Here you can determine a<br />

temperature / current range and<br />

predefine the current input value as<br />

set-point or speed of rotation!<br />

Here you can determine whether<br />

the temperature control unit is<br />

switched OFF/ON via an external<br />

switch contact!<br />

COM.G@TE:<br />

Analogue-Interface<br />

RS232/RS485<br />

ECS Standby<br />

POCO Alarm<br />

Return to main m...<br />

Here you can determine e.g. the<br />

dig. Interface, as well as baudrate<br />

and slave address with RS485!<br />

Here you can determine whether<br />

the unit reports a contact condition<br />

in case of a disturbance (alarm).