F131 Display mode normal<br />
Choosing standard display mode.<br />
F135 Adj. analogue input (analogue interface input current adjustment)<br />
Fine adjustment of the 0/4-20mA or 0-10V input current range. Using this function the<br />
current limits of the 0/4-20mA or 0-10V analogue input signal can be calibrated.<br />
See also chapter on the Com.G@te.<br />
F136 Adj. analogue output (analogue interface output current adjustment)<br />
Fine adjustments of the 0/4-20 mA or 0-10V output current range. Using this function<br />
the current of the unit’s 0/4-20 mA or 0-10V analogue output signal can be calibrated,<br />
this is the current limits for your measured output or difference output via the AIF (see<br />
function F138).<br />
NOTE: Also see chapter on the Com.G@te.<br />
F137 AIF input current / T<br />
Setting the current / temperature assignment for the A / D converter at the analogue<br />
input.<br />
F138 AIF-Output - Function<br />
Assignment of a temperature value to the output current.<br />
F180 Heat. power limit<br />
This function allows the available maximum heating power of the unit to be set<br />
between 0…100%.<br />
F181 Cool. power limit<br />
This function allows the available maximum cooling power of the unit to be set<br />
between 0…100%.<br />