F30 Set date<br />

Setting the date.<br />

F31 Set time<br />

Setting the time.<br />

F33 Set over-temperature protection<br />

Setting the over-temperature protection. Please note chapter on setting the overtemperature<br />

switch.<br />

F34 Air purge<br />

Start / Stop air purge<br />

F35 Compressor automatic<br />

This is used to select the operation of the compressor: Default setting is always ON<br />

Automatic:<br />

The compressor automatic is set to switch on and off as required by the unit.<br />

Benefit: Energy saving<br />

Disadvantage: Longer response times to sudden increase in cooling demand.<br />

Always ON:<br />

The compressor is always running, so the refrigeration system is always immediately<br />

available.<br />

Always OFF:<br />

The compressor is always off.<br />

The compressor automatic (valid for units with compressors only) must be set to<br />

always on when selecting process security in the main menu<br />

Over-temperature protection / OT Mode<br />

F37 Temperature mode<br />

Setting the temperature mode.<br />

F39 Signal<br />

Activating the signal.<br />

F40 Test RS232<br />

Start / Stop<br />

F41 Select user menu<br />

Please refer to chapter select user menu.<br />

F42 Configure user menu<br />

Please refer to chapter configure user menu.<br />

F46 Define analogue input (function analogue input)<br />

This function allows an analogue input current of 0/4-20 mA or 0-10V to be assigned<br />

to an input value.<br />

51