Unipump with echo:<br />
This function is used to monitor if the Unipump being controlled by the POCO is<br />
operating in synchronisation with the Unistat’s own pump. The operating status of the<br />
Unipump can be signalled via a normally open contact by connecting to a “level”<br />
connector. If the Unipump does not operate with the machine, a fault signal will be<br />
generated.<br />
This operating mode is very useful if the Unipump has to be monitored, either to<br />
guarantee the desired temperature control or to avoid unintended heating of the thermal<br />
fluid.<br />
Programmer:<br />
The relay is controlled by a command from a segment within a temperature profile<br />
running on the programmer. Please also note the menu point on Enter program.<br />
Check int. temp. abs. (check internal temperature absolute)<br />
The relay switches when the current internal temperature is outside the specific band<br />
determined by the maximum and minimum temperature limits. Outside this band POCO<br />
is active, within the band POCO is inactive.<br />
Check proc. temp. abs. (check process temperature absolute)<br />
The relay switches when the current external temperature is outside the specific band<br />
determined by the maximum and minimum temperatures. Outside this band POCO is<br />
active, within the band POCO is inactive.<br />