POCO alarm Pot. free Contact (POCO)<br />
Corresponds to Functions F6, F7, and F8. This function allows a relay contact, in the<br />
Com.G@te (46) to be controlled and activated. Please also review the Com.G@te<br />
section of the <strong>manual</strong>.<br />
The following functions are available:<br />
OFF:<br />
The POCO displays the OK status when the unit is ready to operate. This condition<br />
is after the internal controller check, approx. 30 sec. after the switch on. The OK status<br />
will be ended by switching off the mains or a fault.<br />
Check act. value.:<br />
The relay switches when the current internal temperature is outside the range set<br />
between the POCO minimum and maximum internal temperatures (F6 and F7). The<br />
values in functions F6 and F7 are relative to the current set-point. If the range is<br />
exceeded the potential free contact is switched (from the OK status), and the unit will<br />
continue to operate. If the actual value is in the range, the contact will be reset to the<br />
OK status.<br />
External alarm:<br />
The POCO relay is only activated if the unit is in “fault status” when it is switched on.<br />
This is so that the alarm is not raised when the unit gets switched off.<br />
If you wish the alarm function together with the work flow principle, please use the<br />
POCO function OFF.<br />
Unipump / PCS: (Unipump / Process Control System)<br />
This POCO function is used to connect the signal calling for the circulation pump to<br />
start with an external booster pump. This has to be done so that the external pump<br />
runs in synchronisation with the circulation pump in the unit, this means that the POCO<br />
activates (to the OK status) as soon as the circulation pump starts.<br />
PCS: An example of this would be when temperature control would be controlled by a<br />
“PCS" via the external control signal (Menu point External control signal or Function<br />
F28), the POCO can be used to communicate.<br />
Condition POCO ON means temperature control is activated.<br />
Condition POCO OFF means temperature control is not activated.<br />
Control by RS232:<br />
The relay is controlled via an RS232 command. Therefore please note our Huber-<br />
Software.<br />
Check process temperature.:<br />
A measured temperature check for the PROCESS SENSOR providing it is not the control<br />
sensor. The POCO relay switches when the current external, (process temperature) is<br />
outside the range set between the POCO minimum and maximum external<br />
temperatures.<br />
When the unit is set to internal control, and the POCO check process temperature” is<br />
selected, the temperature of the external sensor is monitored – this sensor can be<br />
independent of the internal temperature and the temperature control process. The limits<br />
set by F6 and F7 still apply.<br />
47