POCO alarm Pot. free Contact (POCO)<br />

Corresponds to Functions F6, F7, and F8. This function allows a relay contact, in the<br />

Com.G@te (46) to be controlled and activated. Please also review the Com.G@te<br />

section of the <strong>manual</strong>.<br />

The following functions are available:<br />

OFF:<br />

The POCO displays the OK status when the unit is ready to operate. This condition<br />

is after the internal controller check, approx. 30 sec. after the switch on. The OK status<br />

will be ended by switching off the mains or a fault.<br />

Check act. value.:<br />

The relay switches when the current internal temperature is outside the range set<br />

between the POCO minimum and maximum internal temperatures (F6 and F7). The<br />

values in functions F6 and F7 are relative to the current set-point. If the range is<br />

exceeded the potential free contact is switched (from the OK status), and the unit will<br />

continue to operate. If the actual value is in the range, the contact will be reset to the<br />

OK status.<br />

External alarm:<br />

The POCO relay is only activated if the unit is in “fault status” when it is switched on.<br />

This is so that the alarm is not raised when the unit gets switched off.<br />

If you wish the alarm function together with the work flow principle, please use the<br />

POCO function OFF.<br />

Unipump / PCS: (Unipump / Process Control System)<br />

This POCO function is used to connect the signal calling for the circulation pump to<br />

start with an external booster pump. This has to be done so that the external pump<br />

runs in synchronisation with the circulation pump in the unit, this means that the POCO<br />

activates (to the OK status) as soon as the circulation pump starts.<br />

PCS: An example of this would be when temperature control would be controlled by a<br />

“PCS" via the external control signal (Menu point External control signal or Function<br />

F28), the POCO can be used to communicate.<br />

Condition POCO ON means temperature control is activated.<br />

Condition POCO OFF means temperature control is not activated.<br />

Control by RS232:<br />

The relay is controlled via an RS232 command. Therefore please note our Huber-<br />

Software.<br />

Check process temperature.:<br />

A measured temperature check for the PROCESS SENSOR providing it is not the control<br />

sensor. The POCO relay switches when the current external, (process temperature) is<br />

outside the range set between the POCO minimum and maximum external<br />

temperatures.<br />

When the unit is set to internal control, and the POCO check process temperature” is<br />

selected, the temperature of the external sensor is monitored – this sensor can be<br />

independent of the internal temperature and the temperature control process. The limits<br />

set by F6 and F7 still apply.<br />

47