Protection functions<br />
Following functions are available:<br />
1. Internal sensor high limit alarm<br />
2. Internal sensor low limit alarm<br />
3. Process sensor high limit alarm<br />
4. Process sensor low limit alarm<br />
5. Warning time level (only valid for immersion thermostat <strong>CC</strong>-E and combinations using<br />
the immersion thermostat <strong>CC</strong>-E)<br />
6. Go back<br />
Int. high lim. alarm: (Internal sensor high limit alarm)<br />
Corresponds to the Function F108 in the Function-no. menu.<br />
The temperature monitoring is first activated when the internal (or process) temperature<br />
is below the maximum temperature limit. The temperature must “dip” into the limit<br />
band by 3 K, before an alarm will be triggered. If the temperature limits are below room<br />
temperature, the unit temperature must first reach the temperature band before the<br />
monitoring is activated. This method allows the monitoring temperature to be easily<br />
checked and changed. An alarm is displayed if the temperature value set here is<br />
exceeded for more than 3 seconds.<br />
NOTE: The default setting is set to a value that lies few degrees above the upper<br />
temperature limit of the machine.<br />
Int. low lim. alarm (Internal sensor low limit alarm)<br />
Corresponds to the Function F109 in the Function-no. menu.<br />
An alarm is given when the measured temperature is lower than the set limit values for<br />
more than 3 seconds.<br />
NOTE: The default setting is set to a value that lies few degrees below the lower<br />
temperature limit of the machine.<br />
Proc. high lim. alarm (Process sensor high limit alarm)<br />
Corresponds to the Function nr. F106 in the function menu.<br />
An alarm is displayed if the temperature value set here is exceeded for more than 3<br />
seconds. NOTE: The default setting is set to a value that lies few degrees above the<br />
upper temperature limit of the machine.<br />
Proc. low lim. alarm (Process sensor low limit alarm)<br />
Corresponds to the Function nr. F107 in the function menu.<br />
An alarm is given when the measured temperature is lower than the set limit values for<br />
more than 3 seconds.<br />
NOTE: The default setting is set to a value that lies few degrees below the lower<br />
temperature limit of the machine.<br />