2. With preliminary Test:<br />

To achieve the best control results, select With preliminary Test in the controller<br />

settings. The control parameters will be identified within the limits of the min. and max.<br />

set-points. Temperature control will also take place at the set-point limits. Please be<br />

sure to have set these correctly. Depending on the temperature range, the parameter<br />

search will last correspondingly long.<br />

3. Estimate Control parameters:<br />

In comparison to other temperature control devices on the market which have a fixed<br />

control parameter set, we do however offer an additional feature: an estimated control<br />

parameter, which by selecting the thermal fluid and quantity, then using one of the<br />

parameter sets estimated for your application.<br />

Control dynamics<br />

After the control parameters are detected, control dynamics may be changed (see<br />

example below) without making new identification.<br />

You can select between faster regulation behaviour with a small overshoot (submenu<br />

point Fast, small OS) and a slower regulation behaviour without overshoot (submenu<br />

point Without overshoot). The default setting is “Fast, small OS”. The statement<br />

without overshoot only applies where any interfering action is small.<br />

120<br />

100<br />

80<br />

60<br />

40<br />

20<br />

Without overshoot Adjustment with overshoot<br />

120<br />

set point set point<br />

40<br />

actual value actual value<br />

0<br />

0 5 10 15 20 25 30 35 40 45 50<br />

100<br />

80<br />

60<br />

20<br />

0<br />

0 5 10 15 20 25 30 35 40 45 50<br />

The illustration above shows the transient response for change of set-point.<br />

Note, that you can change the regulation behaviour at any time without having to make<br />

a new controller parameter detection.<br />

Display control parameters<br />

You can have the automatically determined control parameters displayed here.<br />

Conf. <strong>manual</strong> control par. (Configuration <strong>manual</strong> control parameter)<br />

Control parameters can be enetered here. To carry out settings in this mode knowledge<br />

of control technology is required.<br />

Reset control param. (Reset control parameter)<br />

Control parameters can be reset to default setting with this function.<br />