Additional Protection Devices<br />
14<br />
- Auto-Start function<br />
- Alarm function<br />
- Warning messages<br />
- General unit messages<br />
Danger!<br />
Emergency Procedure: Disconnect Electrical Power!<br />
Turn the Mains isolator (36) to “0”!<br />
Dangerous liquid / vapours from temperature control unit or connected hoses (very hot,<br />
very cold, dangerous chemicals) and / or fire / explosion / implosion:<br />
Evacuate the area, following local regulations and procedures to prevent injury or loss<br />
of life! Refer to the MSDS Safety information for the thermal fluid concerned!<br />
Environmental Conditions<br />
This unit, and operations, will comply with DIN EN 61010-1:2001, only when it is<br />
located in suitable environmental conditions.<br />
- for indoor use only;<br />
- installation site ≤ 2000 m altitude;<br />
- installed on a level, even, non flammable surface;<br />
- maintain a clearance above and around the unit of 10 cm for water-cooled units,<br />
and 20cm for air-cooled units, to allow air to circulate around the unit;<br />
- for ambient temperature conditions please refer to the technical data sheet;<br />
remaining within these ambient conditions is imperative in ensuring accurate<br />
operation;<br />
- maximum relative humidity of 80% up to 32°C, decreasing linearly to 50%<br />
relative humidity at 40°C<br />
- use only as long a power cord as necessary;<br />
- the unit should be located so as not to restrict access to the mains power<br />
switch;<br />
- mains voltage should be ±10% of the rated value;<br />
- avoid voltage spikes;<br />
- transient voltage surges as they occur normally in the supply grid;<br />
- clean rating 2;<br />
- overvoltage category II