MANUAL FOLD PLATES<br />
Section 12<br />
12.6 …….Digital Fold Plate - Calibration<br />
1. Remove the lower fold plate cover.<br />
2. The following process is only necessary if the potentiometer has been<br />
replaced - move onto step 3 if this is not the case.<br />
To set up the Potentiometer for the Upper or the Lower Fold plate, see<br />
Fig. 12.15 & Fig. 12.16 and proceed as follows:-<br />
(i) Move the Fold Plate fine adjuster knob so that it is in its central position<br />
(10 mm of thread showing).<br />
(ii) Loosen the gear shown in Fig. 12.15<br />
(iii) By rotating the loose gear perform the following potentiometer adjustment.<br />
Turn the potentiometer fully anti-clockwise and then 5 turns clockwise.<br />
(iv)<br />
(v)<br />
Re-tighten the loose gear.<br />
Move the Fold Plate Deflector or Stop so that it measures 220 mm from the<br />
fold plate nose to the front of the deflector (see Fig. 12.16).<br />
Loosen gear<br />
Fig. 12.15<br />
220 mm<br />
Fig. 12.16<br />
SYSTEMS PAGE 35