30.01.2015 Views

Electromagnetic flow meters - OPTIFLUX 4000 - Forbes Marshall

Electromagnetic flow meters - OPTIFLUX 4000 - Forbes Marshall

Electromagnetic flow meters - OPTIFLUX 4000 - Forbes Marshall

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

2 TECHNICAL DATA<br />

<strong>OPTIFLUX</strong> <strong>4000</strong><br />

Process connections<br />

DIN<br />

ASME<br />

JIS<br />

Design of gasket surface<br />

Electrical connections<br />

Signal cable<br />

Type A<br />

Type B<br />

Approvals and certifications<br />

CE Sign<br />

Hazardous areas<br />

ATEX<br />

FM<br />

CSA<br />

IEC-Ex<br />

NEPSI<br />

DN2.5...3000 in PN 2.5...40 (others on request)<br />

1/10...120" in 150...2500 lbs RF (others on request)<br />

DN2.5...1000 in JIS 10...20 K (others on request)<br />

RF (others on request)<br />

Only for remote systems<br />

Standard cable, double shielded.<br />

Max. length: 600 m / 1950 ft (dep. on electrical conductivity and<br />

measuring sensor). See documentation of the converter for more<br />

information.<br />

Optional cable, triple shielded.<br />

Max. length: 600 m / 1950 ft (dep. on electrical conductivity and<br />

measuring sensor). See documentation of the converter for more<br />

information.<br />

This device fulfills the statutory requirements of the EC directives.<br />

The manufacturer certifies successful testing of the product by<br />

applying the CE mark.<br />

with IFC 100 converter:<br />

KEMA 08 ATEX 0157 X<br />

II 2 G Ex e ia mb IIC T4<br />

II 2 G Ex d ia mb T4<br />

II 2 G Ex e ia mb q T4...T3<br />

II 2 D Ex tD A21 IP64 T120°C<br />

with IFC 300 converter:<br />

KEMA 04 ATEX 2125 X<br />

II 2 GD EEx me ia IIC<br />

II 2 GD EEx de ia IIC<br />

II 2 GD EEx qe ia IIC<br />

II 2 GD EEx e ia IIC<br />

T6...T3 or T5...T3<br />

For more details, see Ex documentation of sensor and converter.<br />

Class I, Div 2, groups A, B, C and D<br />

Class II, Div 2, groups F and G<br />

Class III, Div 2, groups F and G<br />

Class I, Div 2, groups A, B, C and D<br />

Class II, Div 2, groups F and G<br />

pending<br />

GYJ05234 / GYJ05237<br />

Ex me ia IIC T6...T3<br />

Ex de ia II T6...T3<br />

Ex qe ia IIC T6...T3<br />

Ex e ia IIC T6...T3<br />

12<br />

www.krohne.com<br />

03/2009 - <strong>4000</strong>525101 - TD <strong>OPTIFLUX</strong> <strong>4000</strong> R04 en