Calibration 4-19<br />
NECESSARY EQUIPMENT<br />
The following equipment is necessary to complete the adjustments and<br />
calibrations. The suggested equipment or its equivalent may be used.<br />
Instrument Critical Specifications Recommended Model<br />
Analog Oscilloscope 350 MHz Bandwidth Tektronix 2465<br />
Time Interval Counter Frequency Range: 20 MHz min. SRS SR620<br />
Time Interval Accuracy: 1ns max<br />
FFT Spectrum Analyzer Frequency Range: DC to 100 kHz SRS SR760<br />
Amplitude Accuracy: ±0.2 dB<br />
Distortion: < 75 dB below reference<br />
RF Spectrum Analyzer Frequency Range: 1 kHz to 100 MHz Anritsu MS2601/ HP4195A<br />
Amplitude: ±0.5 dB<br />
Distortion and Spurious: < -70 dB<br />
DC/AC Voltmeter 51/2 Digit DC accuracy Fluke 8840A<br />
True RMS AC to 100 kHz<br />
Thermal Converter Input Impedance: 50 Ω Ballantine 1395A-3<br />
Input Voltage: 3 Vrms<br />
Frequency: DC to 10 MHz<br />
Accuracy: ±0.05dB<br />
10 MHz Frequency Standard Frequency: 10 MHz ± .001 ppm SRS FS700<br />
Phase Noise: < -130 dBc @ 100Hz<br />
50 W Terminator 50 Ω ± 0.2 %, 1 Watt HP 11048C<br />
ADJUSTMENTS<br />
The following adjustments set the values of all of the variable components in<br />
the <strong>DS335</strong>. After an adjustment has been made the associated calibrations<br />
must be made. All adjustments must be complete before calibration is<br />
started. First, remove the <strong>DS335</strong>'s top cover by removing the four retaining<br />
screws.) Set the "cal enable" jumper (JP200 switch 2) between pins 1 and 2.<br />
NOTE: The chassis ground and circuit ground float relative to each other.<br />
For voltage measurements use the FUNCTION output BNC shield as a<br />
ground reference.<br />
Output Amplifier Bandwidth<br />
These adjustments correct the bandwidth of the output amplifier. A complete<br />
calibration must be performed if these adjustments are changed. All of the<br />
adjustments are on the bottom PCB. Use an insulated adjusting screwdriver.<br />
1) Connect the output of the <strong>DS335</strong> to the oscilloscope with a 50W<br />
terminator. Set the <strong>DS335</strong> to square wave, 8 Vpp, 100 Hz. Set the<br />
scope to 2 V/div vertical and 5 ms/div horizontal. Adjust R703 for the<br />
squarest output waveform.<br />
2) Set the scope to 200ns/div. Adjust C707 for the fastest output risetime<br />
without excessive overshoots.<br />
3) Do a complete calibration of the <strong>DS335</strong><br />
<strong>DS335</strong> <strong>Synthesized</strong> <strong>Function</strong> <strong>Generator</strong>