30.01.2015 Views

DS335 Synthesized Function Generator

DS335 Synthesized Function Generator

DS335 Synthesized Function Generator

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Calibration 4-17<br />

Now the post amplifier attenuator gains for 50 Ohm case:<br />

40 50_0dB +/- 5% from nominal value of 32768<br />

41 50_2dB +/- 5% from nominal value of 26028.<br />

42 50_4dB +/- 5% from nominal value of 20675.<br />

43 50_6dB +/- 5% from nominal value of 16422.<br />

44 50_8dB +/- 5% from nominal value of 13045.<br />

45 50_10dB +/- 5% from nominal value of 10362.<br />

46 50_12dB +/- 5% from nominal value of 8230.<br />

47 50_14dB +/- 5% from nominal value of 6538.<br />

48 50_16dB +/- 5% from nominal value of 5193.<br />

49 50_18dB +/- 5% from nominal value of 4125.<br />

50 50_20dB +/- 5% from nominal value of 3276.<br />

51 50_22dB +/- 5% from nominal value of 2602.<br />

52 50_24dB +/- 5% from nominal value of 2067.<br />

53 50_26dB +/- 5% from nominal value of 1642.<br />

54 50_28dB +/- 5% from nominal value of 1304.<br />

55 50_30dB +/- 5% from nominal value of 1036.<br />

Spare calbytes (unused):<br />

56 Unused<br />

57 Unused<br />

58 Unused<br />

59 Unused<br />

60 Unused<br />

61 Unused<br />

62 Unused<br />

63 Unused<br />

64 Unused<br />

65 Unused<br />

66 Unused<br />

67 Unused<br />

68 Unused<br />

69 Unused<br />

70 Unused<br />

71 Unused<br />

The following three tables have a length of 160 entries, and contain sine and square wave leveling values and<br />

squarewave symmetry values as a function of frequency.<br />

72 SINE_0 Fixed at 127. This is the reference value for dc.<br />

73 SINE_1 0 to 255. Nominal 127. For 19,531 to 39,062 Hz.<br />

74 SINE_2 0 to 255. Nominal 127. For 39,063 to 58,593 Hz.<br />

:<br />

231 SINE_159 0 to 255. Nominal 127. For 3105468 to 3125000 Hz.<br />

The next table of 160 words contains amplitude leveling data for square wave outputs. This data is used to<br />

modify squarewave amplitude control values as a function of frequency (there is no hardware accommodation<br />

for amplitude leveling of squarewave sweeps.)<br />

232 SQ_0 Fixed at 2^15 (32,768). The dc reference level.<br />

233 SQ_1 For 19,531 to 39,062 Hz. +/-10% from nominal.<br />

234 SQ_2 For 39,063 to 58,593 Hz. +/-10% from nominal.<br />

:<br />

:<br />

391 SQ_159 For 3105468 to 3125000 Hz. +/-10% from nominal.<br />

<strong>DS335</strong> <strong>Synthesized</strong> <strong>Function</strong> <strong>Generator</strong>