Troubleshooting 4-1<br />

TROUBLESHOOTING<br />

If Nothing Happens on<br />

Power On<br />

Make sure that the power entry module on the rear panel is set for the proper<br />

ac line voltage for your location, that the correct fuse is installed, and that the<br />

line cord is inserted all the way into the power entry module. The selected<br />

line voltage may be seen through the clear window, just below the fuse.<br />

When the unit is plugged in and turned "ON", the unit's firmware version<br />

number and serial number will be briefly displayed. Then the self tests should<br />

execute.<br />

Cold Boot<br />

ERROR MESSAGES<br />

Operational Errors<br />

Message<br />

AC Error<br />

AC-DC Error<br />

Freq Error<br />

No GPIB<br />

No RS232<br />

Off Error<br />

Out q err<br />

Range Er<br />

Rate Err<br />

Rcl Err<br />

Span Err<br />

StrtF Er<br />

Stop F Er<br />

If the unit displays no sensible message, the "cold boot" procedure may fix<br />

the problem. To do a "cold boot", turn the unit off. Then, while holding the<br />

"+/-" button, turn the unit "ON". This procedure initializes the RAM and recalls<br />

all factory calibration values.<br />

The following lists explain all of the error messages that the <strong>DS335</strong> can<br />

generate. The messages are divided into operational errors, errors in using<br />

the instrument, and self-test errors. The messages are listed alphabetically.<br />

These error messages may appear during normal front panel operation and<br />

generally are warnings of illegal parameter entries.<br />

Meaning<br />

Amplitude entered is out of allowable range.<br />

The output |V ac | + |V dc | > 5V. Adjust either the offset or amplitude.<br />

Attempt to set output frequency outside of range allowed for current function,<br />

≤ 0 Hz or > 3.1 MHz, or attempting to set frequency for NOISE function.<br />

Cannot access GPIB menus if option board is not installed.<br />

Cannot access RS232 menus if option board is not installed.<br />

DC output offset outside of ±5V range (into 50Ω).<br />

Output queue error. The <strong>DS335</strong> output queue is full due to too many queries<br />

that have not been read back.<br />

Parameter in command is out of allowed range for that command.<br />

Sweep rate out of range (0.01 Hz to 1kHz).<br />

Parameter memory corrupt on power up, stored setting corrupt. Not a worry<br />

unless this error occurs frequently. Check the battery if so.<br />

Logarithmic sweep span error. The start and stop frequencies are apart by<br />

more than six decades.<br />

The sweep start frequency is out of range ( 0 < Freq ≤ max for function).<br />

The sweep stop frequency is out of range ( 0 < Freq ≤ max for function).<br />