Verification of Unmodified FDA- Cleared Test • Verification specimen panel – Own patient specimens • Current test serves as reference method • Split and send to outside lab – Patient specimens from another lab (or vendor) – Old proficiency samples, QC or calibrators • Should be in appropriate matrix, and have analyte in clinically relevant concentrations – Spiked samples (own lab, or provided by vendor) • Appropriate matrix, and analyte in clinically relevant concentrations 10
Verification of Unmodified FDA- Cleared Test • Operators who would perform routine patient testing should perform verification study • Vendors often offer to perform/assist – OK only for fully automated test systems where inter-operator variability not an issue – Preferred assistance: free reagents/kits and data analysis (for complex systems that produce a lot of data, such as AST or serology platform), discrepant analysis 11
- Page 1 and 2: Verification of Microbiology Tests
- Page 3 and 4: Definitions • Modified test: FDA-
- Page 5 and 6: Verification Components • Accurac
- Page 7: Verification of Unmodified FDA- Cle
- Page 13 and 14: Verification of AST • At least 30
- Page 15 and 16: Modified Tests and Laboratory- Deve
- Page 17 and 18: Modification Examples • Change in
- Page 19 and 20: Verification vs. Establishment of
- Page 21 and 22: Verification Report • Summarize p
- Page 23: Acknowledgments • Cumitech 31A co
<strong>Verification</strong> <strong>of</strong> Unmodified FDA-<br />
Cleared Test<br />
• <strong>Verification</strong> specimen panel<br />
– Own patient specimens<br />
• Current test serves as reference method<br />
• Split and send to outside lab<br />
– Patient specimens from another lab (or vendor)<br />
– Old pr<strong>of</strong>iciency samples, QC or calibrators<br />
• Should be in appropriate matrix, and have analyte in<br />
clinically relevant concentrations<br />
– Spiked samples (own lab, or provided by vendor)<br />
• Appropriate matrix, and analyte in clinically relevant<br />
concentrations<br />
10