1.6 Battery Wire Harness<br />

Check that the battery wire harness has the polarity correct.<br />

Set the meter to dc volts and measure the connector with<br />

the red lead on the red wire contact (top of the connector)<br />

and the black lead on the black wire contact (bottom of the<br />

connector) as shown in (figure A1.6.1). If both battery wire<br />

harnesses have full voltage and correct polarity, then proceed<br />

to step 1.8. If voltage is absent proceed to step 1.7. If<br />

polarity is reversed correct battery wiring.<br />

If polarity is reversed correct battery wiring.<br />

A1.6.1<br />

1000 OFF<br />

V DC 200<br />

20<br />

2000m<br />

200m<br />

20M<br />

2000K<br />

200K<br />

20K 2000 200<br />

OHM<br />

750<br />

200V AC<br />

V mA COM 10A<br />

200u<br />

2000u<br />

20m<br />

200m<br />

10A<br />

A DC<br />

DIGITAL MULTIMETER<br />

1.7 Battery Fuse<br />

Check that the battery fuse is in good condition. With the<br />

batteries disconnected remove the fuse cap, inspect the<br />

fuse to see if the fuse is blown. To make sure the fuse is<br />

not blown, set the meter to ohms and measure the resistance<br />

across the fuse. see (figure A1.7.1). If the meter<br />

reads more than one ohm, change the fuse, or else proceed<br />

to the next step.<br />

If the meter reads more than one ohm, change the Battery fuse.<br />

A1.7.1<br />

1000 OFF<br />

V DC 200<br />

20<br />

2000m<br />

200m<br />

20M<br />

2000K<br />

200K<br />

20K 2000 200<br />

OHM<br />

750<br />

200V AC<br />

200u<br />

2000u<br />

20m<br />

200m<br />

10A<br />

V mA COM 10A<br />

A DC<br />

DIGITAL MULTIMETER<br />

1.8 Circuit Breaker Test<br />

To check the circuit breaker set the meter to ohms and<br />

measure the resistance across the circuit breaker as shown<br />

in (figure A1.8.1). If the meter reads more than 1 ohm,<br />

then change the circuit breaker, otherwise proceed to next<br />

step.<br />

A1.8.1<br />

1000 OFF<br />

V DC 200<br />

20<br />

2000m<br />

200m<br />

20M<br />

2000K<br />

200K<br />

20K 2000 200<br />

OHM<br />

750<br />

200V AC<br />

V mA COM 10A<br />

200u<br />

2000u<br />

20m<br />

200m<br />

10A<br />

A DC<br />

DIGITAL MULTIMETER<br />

If the meter reads more than 1 ohm, then change the circuit breaker.<br />

1.9 Main Harness<br />

If the above steps did not correct the problem, change the<br />

main harness.<br />

If the previous steps did not correct the problem, change the main harness.<br />

SUNRISE MEDICAL<br />

QUICKIE S-11 SERVICE MANUAL 2005<br />

PAGE 3