Understanding VSI Controller Diagnostics Codes (cont.)<br />

3.4 Four Bars- Right Motor Disconnected<br />

The right hand motor has a bad connection. Check the connections to the right hand module.<br />

Test Right Motor Open<br />

Check that the batteries are fully charged and in good<br />

condition, and check all cables and connections. Check<br />

the connections to the right motor, look for a loose or damaged<br />

connector. Remove the 9-pin Beau plug and check<br />

the resistance across pin 1 and pin 2 as shown on (figure<br />

A3.4.1). If the meter reads between 0 to 1.5 ohms, then<br />

replace the controller.<br />

1 3 5 7<br />

2 4 6 8<br />

9<br />

A3.4.1<br />

1000 OFF<br />

V DC 200<br />

20<br />

2000m<br />

200m<br />

20M<br />

2000K<br />

200K<br />

20K 2000 200<br />

OHM<br />

750<br />

200V AC<br />

V mA COM 10A<br />

200u<br />

2000u<br />

20m<br />

200m<br />

10A<br />

A DC<br />

DIGITAL MULTIMETER<br />

If the meter reads between 0 and 1.5 ohms, then replace the controller.<br />

Otherwise, check both brushes on the right motor. (Refer to<br />

Section 4 Step 9 of this Service Manual for easiest access).<br />

Ensure that the brushes are not excessively worn, (replace<br />

as required). Top access for Motor brush shown in (figure<br />

A3.4.2) (requires shroud removal).<br />

A3.4.2<br />

Use the meter to check the resistance across the two<br />

bottom contacts of the thicker wires on the 4-pin motor<br />

connector as shown in (figure A3.4.3). If the meter reads<br />

between 0 to 1.5 ohms, then replace the main harness. If<br />

none of the above corrects the problem, replace the right<br />

motor.<br />

1000 OFF<br />

V DC 200<br />

20<br />

2000m<br />

200m<br />

20M<br />

2000K<br />

200K<br />

20K 2000 200<br />

OHM<br />

750<br />

200V AC<br />

200u<br />

2000u<br />

20m<br />

200m<br />

10A<br />

V mA COM 10A<br />

A DC<br />

A3.4.3<br />

DIGITAL MULTIMETER<br />

If the meter reads between 0 and 1.5 ohms, then replace the main harness. If this does not correct the problem,<br />

then replace the right motor.<br />

SUNRISE MEDICAL<br />

QUICKIE S-11 SERVICE MANUAL 2005<br />

PAGE 7