3.1 One Bar - Low Battery Voltage<br />
Section 3<br />
Understanding VSI Controller Diagnostics Codes<br />
This code could indicate discharged batteries, failed batteries, or poor battery connections. Begin by recharging the batteries<br />
and then refer to Section 1 to check batteries and connections.<br />
3.2 Two Bars - Left Motor Disconnected<br />
The left hand motor has a bad connection. Check the connections to the left hand motor.<br />
Test Left Motor Open<br />
Check that the batteries are fully charged and in good<br />
condition; and check all cables and connections. Check<br />
the connections to the left motor, look for a loose or damaged<br />
connector. Remove the 9-pin Beau plug and check<br />
the resistance across pin 3 and pin 4 as shown in (figure<br />
A3.2.1). If the meter reads between 0 to 1.5 ohms, then<br />
replace the controller.<br />
1 3 5 7<br />
2 4 6 8<br />
9<br />
A3.2.1<br />
1000 OFF<br />
V DC 200<br />
20<br />
2000m<br />
200m<br />
20M<br />
2000K<br />
200K<br />
20K 2000 200<br />
OHM<br />
750<br />
200V AC<br />
V mA COM 10A<br />
200u<br />
2000u<br />
20m<br />
200m<br />
10A<br />
A DC<br />
DIGITAL MULTIMETER<br />
If the meter reads between 0 to 1.5 ohms, then replace the controller.<br />
Otherwise, check both brushes on the left motor. (Refer to<br />
Section 4 Step 9 of this Service Manual for easiest access).<br />
Ensure that the brushes are not excessively worn, (replace<br />
as required). Top access for Motor brush shown in (figure<br />
A3.2.2)(requires shroud removal).<br />
A3.2.2<br />
Use the meter to check the resistance across the two bottom<br />
contacts (thicker wires) on the 4-pin motor connector<br />
as shown in (figure A3.2.3). If the meter reads between 0<br />
to 1.5 ohms, then replace the main harness. If none of the<br />
above corrects the problem, replace the left motor.<br />
V DC 1000<br />
200<br />
20<br />
2000m<br />
200m<br />
20M<br />
OFF<br />
750<br />
200V AC<br />
200u<br />
2000u<br />
20m<br />
200m<br />
OHM<br />
2000K<br />
200K<br />
20K 2000 200<br />
10A<br />
A DC<br />
V mA COM 10A<br />
A3.2.3<br />
DIGITAL MULTIMETER<br />
If the meter reads between 0 to 1.5 ohms, then replace the main harness.<br />
If none of the above corrects the problem, replace the left motor.<br />
SUNRISE MEDICAL<br />
QUICKIE S-11 SERVICE MANUAL 2005<br />
PAGE 5