3.1 One Bar - Low Battery Voltage<br />

Section 3<br />

Understanding VSI Controller Diagnostics Codes<br />

This code could indicate discharged batteries, failed batteries, or poor battery connections. Begin by recharging the batteries<br />

and then refer to Section 1 to check batteries and connections.<br />

3.2 Two Bars - Left Motor Disconnected<br />

The left hand motor has a bad connection. Check the connections to the left hand motor.<br />

Test Left Motor Open<br />

Check that the batteries are fully charged and in good<br />

condition; and check all cables and connections. Check<br />

the connections to the left motor, look for a loose or damaged<br />

connector. Remove the 9-pin Beau plug and check<br />

the resistance across pin 3 and pin 4 as shown in (figure<br />

A3.2.1). If the meter reads between 0 to 1.5 ohms, then<br />

replace the controller.<br />

1 3 5 7<br />

2 4 6 8<br />

9<br />

A3.2.1<br />

1000 OFF<br />

V DC 200<br />

20<br />

2000m<br />

200m<br />

20M<br />

2000K<br />

200K<br />

20K 2000 200<br />

OHM<br />

750<br />

200V AC<br />

V mA COM 10A<br />

200u<br />

2000u<br />

20m<br />

200m<br />

10A<br />

A DC<br />

DIGITAL MULTIMETER<br />

If the meter reads between 0 to 1.5 ohms, then replace the controller.<br />

Otherwise, check both brushes on the left motor. (Refer to<br />

Section 4 Step 9 of this Service Manual for easiest access).<br />

Ensure that the brushes are not excessively worn, (replace<br />

as required). Top access for Motor brush shown in (figure<br />

A3.2.2)(requires shroud removal).<br />

A3.2.2<br />

Use the meter to check the resistance across the two bottom<br />

contacts (thicker wires) on the 4-pin motor connector<br />

as shown in (figure A3.2.3). If the meter reads between 0<br />

to 1.5 ohms, then replace the main harness. If none of the<br />

above corrects the problem, replace the left motor.<br />

V DC 1000<br />

200<br />

20<br />

2000m<br />

200m<br />

20M<br />

OFF<br />

750<br />

200V AC<br />

200u<br />

2000u<br />

20m<br />

200m<br />

OHM<br />

2000K<br />

200K<br />

20K 2000 200<br />

10A<br />

A DC<br />

V mA COM 10A<br />

A3.2.3<br />

DIGITAL MULTIMETER<br />

If the meter reads between 0 to 1.5 ohms, then replace the main harness.<br />

If none of the above corrects the problem, replace the left motor.<br />

SUNRISE MEDICAL<br />

QUICKIE S-11 SERVICE MANUAL 2005<br />