1.2. Specifications<br />
Accuracy..........................................................<strong>HFM</strong>-<strong>301</strong>/<strong>305</strong> & <strong>HFC</strong>-<strong>303</strong>/<strong>307</strong>: ±1% full scale (F.S.)<br />
...........................................................................................................<strong>HFM</strong>-306; ±3% full scale (F.S.)<br />
Repeatability ................................................................................................................ ±0.07% of F.S.<br />
Maximum pressure................................................................................................500 psi [3.45 MPa]<br />
..............................................................................................................................300 psi (<strong>HFM</strong>-306)<br />
Maximum pressure................................................................................................1000 psi [6.9 MPa]<br />
..................................................................... (with high pressure option on <strong>HFM</strong>-<strong>301</strong>/<strong>HFC</strong><strong>303</strong> only)<br />
Pressure coefficient.................................................................................