3.8. Electronic Circuitry<br />
The 300 Series flow meters employs a thermal transfer principle (capillary tube described in section 3.2)<br />
to measure the flow through the sensor which is proportional to the total flow through the instrument.<br />
The sensor develops a differential voltage output signal proportional to flow, which is amplified to<br />
produce 5 VDC at full scale flow. The amplified output can be measured on pins 6 and 1 of the external<br />
“D” connector. If a <strong>Hastings</strong> power supply is employed, the 5 Volt output is also sent to the terminals on<br />
the back and to the decoding circuitry in the display, which converts it to a numeric output. The<br />
optional 4 - 20 mA analog output is available in lieu of an output voltage. The addition of a 4 - 20 mA<br />
current loop transmitter on a secondary PC board (mounted parallel to the main pc board) is required to<br />
provide this current loop. A jumper change is made on the secondary PC board to establish the selected<br />
output mode.<br />
3.9. Instrument Performance<br />
The combination of these principles embodied in the 300 Series give inherently linear and highly<br />
accurate output and the typical response time from zero to full flow is also quite good. The meter<br />
settling time of a 1000 slm full scale flow meter is typically under 0.5 seconds. Flow controller response<br />
is less than 2 seconds.<br />
Manual: 152-082010 <strong>301</strong>-<strong>305</strong>-306_<strong>303</strong>-<strong>307</strong> Series Page 24 of 35