<strong>Stopwatch</strong> <strong>and</strong> <strong>Timer</strong> <strong>Calibrations</strong><br />
prefixes that can be used with seconds, as well as the multiples/submultiples <strong>and</strong><br />
symbols used to represent them. The second is one <strong>of</strong> the seven base units in the<br />
International System <strong>of</strong> Units (SI). Other units (most notably the meter <strong>and</strong> the<br />
volt) have definitions that depend upon the definition <strong>of</strong> the second.<br />
Table 1 - Metric prefixes<br />
(may be applied to all SI units)<br />
Multiples <strong>and</strong> Submultiples Prefix Symbol<br />
1 000 000 000 000 = 10 12 tera T<br />
1 000 000 000 = 10 9 giga G<br />
1 000 000 = 10 6 mega M<br />
1 000 = 10 3 kilo k<br />
1 = 10 0<br />
0.001 = 10 -3 milli m<br />
0.000 001 = 10 -6 micro <br />
0.000 000 001 = 10 -9 nano n<br />
0.000 000 000 001 = 10 -12 pico p<br />
0.000 000 000 000 001 = 10 -15 femto f<br />
The SI defines the second based on a property <strong>of</strong> the cesium atom, <strong>and</strong> for this<br />
reason cesium oscillators are regarded as primary st<strong>and</strong>ards for both time interval<br />
<strong>and</strong> frequency. A second is defined as the time interval required for 9 192 631 770<br />
transitions between two energy states <strong>of</strong> the cesium atom to take place. The atomic<br />
definition <strong>of</strong> the second, together with current technology, allows it to be measured<br />
with much smaller uncertainties than any other SI unit. In fact, the <strong>National</strong><br />
<strong>Institute</strong> <strong>of</strong> St<strong>and</strong>ards <strong>and</strong> Technology (NIST) can currently measure a second<br />
with an uncertainty <strong>of</strong> less than 1 part in 10 15 , or more than 1 billion (10 9 ) times<br />
smaller than the uncertainties required for the calibrations described in this guide!<br />
The resolution <strong>of</strong> a stopwatch or timer represents the smallest time interval<br />
that the device can display. Resolution is related to the number <strong>of</strong> digits on the<br />
device’s display for a digital stopwatch, or the smallest increment or graduation<br />
on the face <strong>of</strong> an analog stopwatch. For example, if a stopwatch display shows<br />
two digits to the right <strong>of</strong> the decimal point, it has a resolution <strong>of</strong> 0.01 s (10 ms, or<br />
1/100 <strong>of</strong> a second). This means, for example, that it can display a value <strong>of</strong> 42.12<br />
2