CHAPTER 9. SUMMARY 56<br />
signal-to-noise ratios were observed to be 46 and 7.2, respectively, achieving approximately<br />
digital spatial resolution in both cases. However, the version of the SCT chip was not yet<br />
optimized <strong>for</strong> the small capacitive load of diamond detectors. The spatial resolution seems<br />
to be limited by the polycrystalline structure of CVD diamond. Furthermore, the rst<br />
prototype of a diamond pixel detector (ATLAS design) demonstrated its functionality in<br />
atestbeam, and digital spatial resolution was observed in both dimensions.<br />
The future program of RD42 includes further improvement of the charge collection<br />
distance as well as the growing of large area detectors. Emphasis will be laid upon<br />
the preparation and test of pixel detectors. Furthermore, the homogeneity studies will<br />
be continued to investigate the charge collection properties on a scale of a few tens of<br />
micrometers in the lateral dimension.