Cable and Antenna Analyzer (continued) Description Specification Supplemental Information Distance–to–Fault Horizontal Range Range = [(number of points – 1) / frequency span * 2] * velocity factor * speed of light Horizontal Resolution Bandpass Mode Window Types 1 Using recommended calibration kits. Resolution = Range / (number of points – 1) Number of points auto coupled according to start and stop distance entered Number of points settable by user Maximum, medium, and minimum windows 2 Higher cable losses can be measured using transmission or S21 measurements. Cable losses measured in transmission mode limited by transmission dynamic range. 3 QuickCal is performed with the connect LOAD step. Figure 1: CAT Mode, Type–N Calibration Kit – Magnitude (Specification) 6

Cable and Antenna Analyzer (continued) Figure 2: CAT Mode, QuickCal – Magnitude (Typical) Figure 3: CAT Mode, Preset Cal – Magnitude (Typical) 7

Cable and Antenna <strong>Analyzer</strong> (continued)<br />

Description Specification Supplemental Information<br />

Distance–to–Fault<br />

Horizontal Range Range = [(number of points – 1) /<br />

frequency span * 2] * velocity factor *<br />

speed of light<br />

Horizontal<br />

Resolution<br />

Bandpass Mode<br />

Window Types<br />

1 Using recommended calibration kits.<br />

Resolution = Range / (number of<br />

points – 1)<br />

Number of points auto coupled according to start<br />

and stop distance entered<br />

Number of points settable by user<br />

Maximum, medium, and minimum windows<br />

2 Higher cable losses can be measured using transmission or S21 measurements. Cable losses measured in<br />

transmission mode limited by transmission dynamic range.<br />

3 QuickCal is performed with the connect LOAD step.<br />

Figure 1: CAT Mode, Type–N Calibration Kit – Magnitude (Specification)<br />