Agilent FieldFox RF Analyzer N9912A - Tessco
Agilent FieldFox RF Analyzer N9912A - Tessco Agilent FieldFox RF Analyzer N9912A - Tessco
Cable and Antenna Analyzer (continued) Description Specification Supplemental Information Distance–to–Fault Horizontal Range Range = [(number of points – 1) / frequency span * 2] * velocity factor * speed of light Horizontal Resolution Bandpass Mode Window Types 1 Using recommended calibration kits. Resolution = Range / (number of points – 1) Number of points auto coupled according to start and stop distance entered Number of points settable by user Maximum, medium, and minimum windows 2 Higher cable losses can be measured using transmission or S21 measurements. Cable losses measured in transmission mode limited by transmission dynamic range. 3 QuickCal is performed with the connect LOAD step. Figure 1: CAT Mode, Type–N Calibration Kit – Magnitude (Specification) 6
Cable and Antenna Analyzer (continued) Figure 2: CAT Mode, QuickCal – Magnitude (Typical) Figure 3: CAT Mode, Preset Cal – Magnitude (Typical) 7
- Page 1 and 2: Agilent FieldFox RF Analyzer N9912A
- Page 3 and 4: Table of Contents Definitions Speci
- Page 5: Maximum Measurable Cable Loss Using
- Page 9 and 10: Display data Trace math Scale Title
- Page 11 and 12: Network Analyzer (continued) Figure
- Page 13 and 14: Spectrum Analyzer (Option 230 and 2
- Page 15 and 16: Spectrum Analyzer (continued) Descr
- Page 17 and 18: Spectrum Analyzer (continued) Descr
- Page 19 and 20: Spectrum Analyzer (continued) Descr
- Page 21 and 22: Receiver IF Bandwidth AM FM Narrow
- Page 23 and 24: General Information (continued) Des
- Page 25 and 26: Supported Cal Kits The following li
Cable and Antenna <strong>Analyzer</strong> (continued)<br />
Description Specification Supplemental Information<br />
Distance–to–Fault<br />
Horizontal Range Range = [(number of points – 1) /<br />
frequency span * 2] * velocity factor *<br />
speed of light<br />
Horizontal<br />
Resolution<br />
Bandpass Mode<br />
Window Types<br />
1 Using recommended calibration kits.<br />
Resolution = Range / (number of<br />
points – 1)<br />
Number of points auto coupled according to start<br />
and stop distance entered<br />
Number of points settable by user<br />
Maximum, medium, and minimum windows<br />
2 Higher cable losses can be measured using transmission or S21 measurements. Cable losses measured in<br />
transmission mode limited by transmission dynamic range.<br />
3 QuickCal is performed with the connect LOAD step.<br />
Figure 1: CAT Mode, Type–N Calibration Kit – Magnitude (Specification)<br />
6