Maximum Measurable Cable Loss Using 1–Port CAT Measurement Model 2 Refl Dyn Range /2<br />

Cable and Antenna <strong>Analyzer</strong> (continued)<br />

Description Specification Typical<br />

10 minute warm up 90 minute warm up<br />

Directivity<br />

Corrected with OSL calibration 1 >42 dB >42 dB<br />

Corrected with QuickCal (Option 111) 3<br />

≥42 dB<br />

Raw<br />

2 MHz to 3.5 GHz > 20 dB<br />

> 3.5 GHz to 6 GHz > 14 dB<br />

Source Match<br />

Corrected with OSL calibration 1 > 36 dB > 36 dB<br />

Corrected with QuickCal (Option 111) 3<br />

≥35 dB<br />

Raw<br />

2 MHz to 3 GHz > 25 dB<br />

> 3 GHz to 6 GHz > 16 dB<br />

Reflection Tracking<br />

Corrected with OSL calibration 1 ±0.06 dB ±0.06 dB<br />

Corrected with QuickCal (Option 111) 3<br />

±0.15 dB<br />

Reflection Dynamic Range<br />

Reflection (<strong>RF</strong> Out port) (High power out)<br />

2 MHz to 4 GHz 60 dB<br />

> 4 GHz to 6 GHz 55 dB<br />

Transmission Dynamic Range(Option 110)<br />

300 Hz IF Bandwidth<br />

2 MHz to 2 GHz 72 dB<br />

> 2 GHz to 3 GHz 67 dB<br />

> 3 GHz to 5 GHz 58 dB<br />

> 5 GHz to 6 GHz 49 dB<br />

Return Loss<br />

Display Range<br />

Resolution<br />

0 to 100 dB<br />

0.01 dB<br />

VSWR<br />

Display Range 1 to 500<br />

Resolution 0.01<br />

Cable Loss<br />

Display Range<br />

Resolution<br />

0 to 100 dB<br />

0.01 dB<br />