20.01.2015 Views

View - ResearchGate

View - ResearchGate

View - ResearchGate

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

310 Heegaard et al.<br />

characteristics may be desired, e.g. when studying metal-ion-binding proteins<br />

(40,41). Calcium ions will, for example, precipitate in phosphate buffers. Very<br />

reliable results may instead be obtained with HEPES buffers that have minimal<br />

cation-binding (42). In work involving, for example Ca 2+ , it may be necessary<br />

to use chelating agents such as ethylenediaminetetraacetic acid (EDTA) in the<br />

washing solutions to remove all divalent cations between runs. The magnitude<br />

of the EEO flow will be a sensitive indicator of the amount of immobilized<br />

cations in such experiments.<br />

The interplay between sample solution and electrophoresis buffer also<br />

requires attention. Conductivity differences may be detrimental but may also<br />

be exploited to increase detection limits by taking advantage of stacking<br />

phenomena. It is important to realize that even though considerable increases<br />

in detection limits may be achieved by dissolving the analyte in a sample<br />

buffer with lower (typically 1/10 diluted electrophoresis buffer) conductivity<br />

than the electrophoresis buffer (43,44), the resulting temperature increase in the<br />

sample zone may be very high leading to, for example, heat-induced partial or<br />

complete denaturation or the induction of other artifacts, such as, aggregation<br />

of the protein analyte (22) (see Fig. 2). Conversely, when the conductivity of<br />

the sample is higher (e.g. because of a high salt content), analyte peak broadening<br />

is to be expected. Also, in any CE experiment where buffer and sample<br />

conductivity is not the same, the precise concentration of the analyte in the<br />

sample zone is bound to be different from the concentration in the sample and<br />

will change during the initial electrophoresis steps. This complicates affinity<br />

experiments where the exact concentration of analyte during the run is required<br />

for the subsequent calculations.<br />

In addition to conductivity differences, the correspondence of pH in the<br />

sample solution and in the electrophoresis buffer also warrants attention because<br />

the crossing of the pH boundary created upon initiation of electrophoresis may<br />

lead to analyte aggregation and precipitation.<br />

Finally, the vial strategy should be considered for two main reasons: one is<br />

that repeated electrophoresis from the same buffer vial will lead to so-called<br />

buffer depletion, (a change, caused by electrolysis, in the ionic composition of<br />

the anodic and cathodic buffer solutions) leading to changes in mobility when<br />

the electrolyzed buffer is used as a running buffer. Thus, fresh buffer should<br />

always be used to replenish the electrophoresis buffer, for example, by using<br />

different reservoirs for running and for rinsing. This will ensure a reproducible<br />

composition of the buffer inside the capillary. Another detail regarding vial and<br />

washing strategies is that in affinity electrophoresis with ligand addition to the<br />

electrophoresis buffer, it is normally not the intention to introduce ligand into<br />

the sample solution. Carry over of ligand into the sample solution when sample<br />

injection follows immediately after washing the capillary with the ligand-

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!