260 Benčina et al.<br />

3.1.1. Static Immobilization Method<br />

1. Place CIM disk into CIM housing to obtain CIM disk monolithic column.<br />

2. Connect CIM disk monolithic column to HPLC system and equilibrate it by<br />

washing with at least 5 column volumes of water and at least 5 column volumes<br />

of immobilization buffer (see Note 1).<br />

3. Prepare 3 ml of a protein solution (2 g/l) by dissolving the protein in appropriate<br />

immobilization buffer.<br />

4. Remove CIM disk from the CIM housing and immerse it into 3 ml of the immobilization<br />

solution (see Fig. 1A).<br />

5. Incubate CIM disk in the immobilization solution at given temperature for a<br />

defined period of time (see Note 1).<br />

6. After immobilization is completed, place CIM disk again into CIM housing,<br />

connect CIM disk monolithic column (named further “enzyme reactor”) to HPLC<br />

system, remove the residual non-bound protein by washing the enzyme reactor<br />

with 10 column volumes of immobilization buffer containing 0.1 M NaCl and<br />

finally with deionized water.<br />

7. Disconnect enzyme reactor from HPLC system, remove CIM disk from CIM<br />

housing and store immobilized CIM disk at 4°C either in water, 20% ethanol, or<br />

suitable buffer.<br />

3.1.2. Dynamic Immobilization Method<br />

1. Place CIM disk into CIM housing to obtain CIM disk monolithic column.<br />

2. Connect a syringe filled with water to one side of the CIM disk monolithic column.<br />

3. Equilibrate CIM disk monolithic column by pushing with a syringe at least 5<br />

column volumes of water (∼2 ml), fill the syringe with immobilization buffer and<br />

wash the column with at least 5 column volumes.<br />

4. Prepare 3 ml of a protein solution (2 g/l) by dissolving the protein in appropriate<br />

buffer and fill the syringe with it.<br />

Fig. 1. Static (A) and dynamic (B) immobilization method.