CENTRA-S 200, R 200<br />

CENTRA-S/R<br />

• Centralized purification and distribution<br />

systems capable of supplying pure water to<br />

one or more laboratories on a loop system<br />

• Both models feature a 200 liters per hour<br />

reverse osmosis module. <strong>The</strong> CENTRA-R<br />

additionally includes UV photo-oxidation and<br />

a feed to and from a deionization cylinder<br />

• Incorporates a wide range of safety features<br />

and alarms to ensure pure water quality is<br />

never compromised<br />

• Standard CENTRA systems deliver a loop<br />

flowrate of up to 18 liters per minute.<br />

Additionally the CENTRA-S 200 and R 200 are<br />

available as high flow variable speed models,<br />

(HFV) capable of delivering up to 30 liters<br />

per minute<br />

• Compatible with building management<br />

systems<br />

Standard CENTRA Model S 200 R 200<br />

Loop Flow rate Up to 18 liters per minute Up to 18 liters per minute<br />

High Flow CENTRA Model S 200-HFV R 200-HFV<br />

Typical Loop Flow Rate Variable speeds, up to a maximum of 30 liters per minute 1<br />

Reservoir Capacity 350 liters 350 liters<br />

Recirculation Pump √ √<br />

Reverse Osmosis √ √<br />

UV Photo-oxidation<br />

0.2μm Filter √ √<br />

Optional Deionization<br />

Product <strong>Water</strong> Specifications<br />

<strong>The</strong> performance specifications given below are intended only as a guide.<br />

Each system can be individually specified to meet your particular requirements.<br />

Inorganics (resistivity @ 25°C) RO permeate Up to >18.2 MΩ-cm 3<br />