Proline Promass 83E/F/O Description of Device Functions - FMC ...
Proline Promass 83E/F/O Description of Device Functions - FMC ...
Proline Promass 83E/F/O Description of Device Functions - FMC ...
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />
7 Block OUTPUTS<br />
Function description<br />
OUTPUTS PULSE/FREQUENCY OUTPUT (1...2) OPERATION (STATUS)<br />
ACTUAL STATUS<br />
(4341) ! Note!<br />
This function is not available unless the STATUS setting was selected in the OPERATION<br />
MODE function (4200).<br />
Use this function to check the current status <strong>of</strong> the status output.<br />
Display:<br />
NOT CONDUCTIVE<br />
CONDUCTIVE<br />
SIMULATION SWITCH<br />
POINT<br />
(4342)<br />
! Note!<br />
This function is not available unless the STATUS setting was selected in the OPERATION<br />
MODE function (4200).<br />
Use this function to activate simulation <strong>of</strong> the status output.<br />
Options:<br />
OFF<br />
ON<br />
Factory setting:<br />
OFF<br />
! Note!<br />
• The “SIMULATION STATUS OUTPUT” message indicates that simulation is active.<br />
• The measuring device continues to measure while simulation is in progress, i.e. the<br />
current measured values are output correctly via the other outputs.<br />
" Caution!<br />
The setting is not saved if the power supply fails.<br />
VALUE SIMULATION<br />
SWITCH POINT<br />
(4343)<br />
! Note!<br />
This function is not available unless STATUS was selected in the OPERATION MODE<br />
function (4200) and the SIMULATION SWITCH POINT function (4342) is active<br />
(= ON).<br />
Use this function to define the switching response <strong>of</strong> the status output during the<br />
simulation. This value is used to test downstream devices and the measuring device itself.<br />
Options:<br />
NOT CONDUCTIVE<br />
CONDUCTIVE<br />
Factory setting:<br />
NOT CONDUCTIVE<br />
" Caution!<br />
The setting is not saved if the power supply fails.<br />
Issue/Rev. 0.2 (12/12) MN0M023 • Page 99