7 Block OUTPUTS <strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />
Function description<br />
OUTPUTS PULSE/FREQUENCY OUTPUT (1...2) OPERATION (PULSE)<br />
SIMULATION<br />
PULSE(4322) ! Note!<br />
This function is not available unless the PULSE option was selected in the OPERATION<br />
MODE function.<br />
Use this function to activate simulation <strong>of</strong> the pulse output.<br />
Options:<br />
OFF<br />
COUNTDOWN<br />
The pulses specified in the VALUE SIMULATION PULSE function are output.<br />
CONTINUOUSLY<br />
Pulses are continuously output with the pulse width specified in the PULSE WIDTH<br />
function. Simulation is started once the CONTINUOUSLY option is confirmed with the<br />
F key.<br />
! Note!<br />
Simulation is started by confirming the CONTINUOUSLY option with the F key. The<br />
simulation can be switched <strong>of</strong>f again via the SIMULATION PULSE function.<br />
Factory setting:<br />
OFF<br />
! Note!<br />
• The notice message #631 “SIM. PULSE” indicates that simulation is active.<br />
• The on/<strong>of</strong>f ratio is 1:1 for both types <strong>of</strong> simulation.<br />
• The measuring device continues to measure while simulation is in progress, i.e. the<br />
current measured values are output correctly via the other outputs.<br />
" Caution!<br />
The setting is not saved if the power supply fails.<br />
VALUE SIMULATION<br />
PULSE<br />
(4323)<br />
! Note!<br />
This function is not available unless the COUNTDOWN option was selected in the<br />
SIMULATION PULSE function.<br />
Use this function to specify the number <strong>of</strong> pulses (e.g. 50) which are output during the<br />
simulation. This value is used to test downstream devices and the measuring device itself.<br />
The pulses are output with the pulse width specified in the PULSE WIDTH function. The<br />
on/<strong>of</strong>f ratio is 1:1.<br />
Simulation is started once the specified value is confirmed with the F key. The display<br />
remains at 0 if the specified pulses have been output.<br />
User input:<br />
0...10 000<br />
Factory setting:<br />
0<br />
! Note!<br />
Simulation is started by confirming the simulation value with the F key. The simulation<br />
can be switched <strong>of</strong>f again via the SIMULATION PULSE function.<br />
" Caution!<br />
The setting is not saved if the power supply fails.<br />
Page 98 • MN0M023<br />