7 Block OUTPUTS <strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />

Function description<br />

OUTPUTS PULSE/FREQUENCY OUTPUT (1...2) OPERATION (PULSE)<br />

SIMULATION<br />

PULSE(4322) ! Note!<br />

This function is not available unless the PULSE option was selected in the OPERATION<br />

MODE function.<br />

Use this function to activate simulation <strong>of</strong> the pulse output.<br />

Options:<br />

OFF<br />

COUNTDOWN<br />

The pulses specified in the VALUE SIMULATION PULSE function are output.<br />

CONTINUOUSLY<br />

Pulses are continuously output with the pulse width specified in the PULSE WIDTH<br />

function. Simulation is started once the CONTINUOUSLY option is confirmed with the<br />

F key.<br />

! Note!<br />

Simulation is started by confirming the CONTINUOUSLY option with the F key. The<br />

simulation can be switched <strong>of</strong>f again via the SIMULATION PULSE function.<br />

Factory setting:<br />

OFF<br />

! Note!<br />

• The notice message #631 “SIM. PULSE” indicates that simulation is active.<br />

• The on/<strong>of</strong>f ratio is 1:1 for both types <strong>of</strong> simulation.<br />

• The measuring device continues to measure while simulation is in progress, i.e. the<br />

current measured values are output correctly via the other outputs.<br />

" Caution!<br />

The setting is not saved if the power supply fails.<br />

VALUE SIMULATION<br />

PULSE<br />

(4323)<br />

! Note!<br />

This function is not available unless the COUNTDOWN option was selected in the<br />

SIMULATION PULSE function.<br />

Use this function to specify the number <strong>of</strong> pulses (e.g. 50) which are output during the<br />

simulation. This value is used to test downstream devices and the measuring device itself.<br />

The pulses are output with the pulse width specified in the PULSE WIDTH function. The<br />

on/<strong>of</strong>f ratio is 1:1.<br />

Simulation is started once the specified value is confirmed with the F key. The display<br />

remains at 0 if the specified pulses have been output.<br />

User input:<br />

0...10 000<br />

Factory setting:<br />

0<br />

! Note!<br />

Simulation is started by confirming the simulation value with the F key. The simulation<br />

can be switched <strong>of</strong>f again via the SIMULATION PULSE function.<br />

" Caution!<br />

The setting is not saved if the power supply fails.<br />

Page 98 • MN0M023<br />

Issue/Rev. 0.2 (12/12)