<strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />

7 Block OUTPUTS<br />

Function description<br />

OUTPUTS PULSE/FREQUENCY OUTPUT (1...2) OPERATION (FREQUENCY)<br />

VALUE SIMULATION<br />

FREQUENCY<br />

(4303)<br />

! Note!<br />

This function is not available unless FREQUENCY was selected in the OPERATION<br />

MODE function (4200) and the SIMULATION FREQUENCY function (4302) is active<br />

(= ON).<br />

Use this function to define a free selectable frequency value (e.g. 500 Hz) which should<br />

be output at the frequency output (with maximum pulse frequency or shortened<br />

minimum pulse width). This value is used to test downstream devices and the measuring<br />

device itself.<br />

User input:<br />

0...12500 Hz<br />

Factory setting:<br />

0 Hz<br />

" Caution!<br />

The setting is not saved if the power supply fails.<br />