Proline Promass 83E/F/O Description of Device Functions - FMC ...
Proline Promass 83E/F/O Description of Device Functions - FMC ...
Proline Promass 83E/F/O Description of Device Functions - FMC ...
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
7 Block OUTPUTS <strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />
7.2.2 Function group OPERATION<br />
OUTPUTS E CURRENT OUTPUT 1 EAA<br />
<br />
CURRENT OUTPUT 2 EAB<br />
<br />
CURRENT OUTPUT 3 EAC<br />
<br />
PULSE/FREQ. OUTPUT 1 ECA CONFIGURATION 420<br />
<br />
OPERATION 430<br />
PULSE/FREQ. OUTPUT 2 ECB CONFIGURATION 420<br />
<br />
OPERATION 430<br />
Function description<br />
OUTPUTS PULSE/FREQUENCY OUTPUT (1...2) OPERATION (FREQUENCY)<br />
ACTUAL FREQUENCY<br />
(4301) ! Note!<br />
This function is not available unless the FREQUENCY setting was selected in the<br />
OPERATION MODE function (4200).<br />
Use this function to view the computed actual value <strong>of</strong> the output frequency.<br />
Display:<br />
0...12500 Hz<br />
SIMULATION<br />
FREQUENCY<br />
(4302)<br />
! Note!<br />
This function is not available unless the FREQUENCY setting was selected in the<br />
OPERATION MODE function (4200).<br />
Use this function to activate simulation <strong>of</strong> the frequency output.<br />
Options:<br />
OFF<br />
ON<br />
Factory setting:<br />
OFF<br />
! Note!<br />
• The “SIMULATION FREQUENCY OUTPUT” message indicates that simulation is<br />
active.<br />
• The measuring device continues to measure while simulation is in progress, i.e. the<br />
current measured values are output correctly via the other outputs.<br />
" Caution!<br />
The setting is not saved if the power supply fails.<br />
Page 96 • MN0M023<br />
Issue/Rev. 0.2 (12/12)