<strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />

7 Block OUTPUTS<br />

Function description<br />

OUTPUTS PULSE/FREQUENCY OUTPUT (1...2) CONFIGURATION (PULSE)<br />

FAILSAFE MODE<br />

(4227) ! Note!<br />

This function is not available unless the PULSE setting was selected in the OPERATION<br />

MODE function (4200).<br />

For safety reasons it is advisable to ensure that the pulse output assumes a predefined<br />

state in the event <strong>of</strong> a fault. The setting you select here affects only the pulse output. It<br />

has no effect on other outputs and the display (e.g. totalizers).<br />

Options:<br />

FALLBACK VALUE<br />

Output is 0 pulse.<br />

ACTUAL VALUE<br />

Measured value output is based on the current flow measurement. The fault is ignored.<br />

Factory setting:<br />

FALLBACK VALUE<br />