7 Block OUTPUTS <strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />
Function description<br />
OUTPUTS PULSE/FREQUENCY OUTPUT (1...2) CONFIGURATION (PULSE)<br />
OUTPUT SIGNAL<br />
(continued)<br />
Example for active output circuit (ACTIVE):<br />
With an active circuit, the internal power supply is 24 V.<br />
The pulse output is short-circuit pro<strong>of</strong>.<br />
=<br />
-<br />
➀<br />
+<br />
➁<br />
1 5 7 8 3<br />
m = 24 V DC internal power supply<br />
n = Short-circuit pro<strong>of</strong> output<br />
A0004691<br />
The signal levels are to be seen as analogous to the passive circuit.<br />
The following applies for the output configuration ACTIVE-POSITIVE:<br />
In the quiescent state (at zero flow), the output signal level at the terminals is 0 V.<br />
U<br />
24 V<br />
0V<br />
t<br />
A0004694<br />
In the operating status (flow present), the output signal level changes from 0 V to a<br />
positive voltage level.<br />
24 V<br />
U<br />
0V<br />
t<br />
A0004692<br />
The following applies for the output configuration ACTIVE-NEGATIVE:<br />
In the quiescent state (at zero flow), the output signal level at the terminals is at a positive<br />
voltage level.<br />
24 V<br />
U<br />
0V<br />
t<br />
A0004693<br />
In the operating status (flow present), the output signal level changes from a positive<br />
voltage level to 0 V.<br />
24 V<br />
U<br />
0V<br />
t<br />
A0004710<br />