<strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />

7 Block OUTPUTS<br />

Function description<br />

OUTPUTS PULSE/FREQUENCY OUTPUT (1...2) CONFIGURATION (PULSE)<br />

MEASURING MODE<br />

(4225) ! Note!<br />

This function is not available unless the PULSE setting was selected in the OPERATION<br />

MODE function (4200).<br />

Use this function to define the measuring mode for the pulse output.<br />

Options:<br />

STANDARD<br />

Only positive flow components are totaled. Negative components are not taken into<br />

account.<br />

SYMMETRY<br />

Positive and negative flow components are taken into account.<br />

! Note!<br />

The direction <strong>of</strong> flow can be output via the relay output.<br />

PULSATING FLOW<br />

If flow is characterized by severe fluctuations as is the case, for example, with<br />

reciprocating pumps, the positive and negative flow components are totaled, with the<br />

signs taken into account (e.g. -10 l and +25 l = 15 l).<br />

Flow components outside the maximum pulse number per second (value/width) are<br />

buffered, balanced and output after a maximum delay <strong>of</strong> 60 seconds. If the buffered data<br />

cannot be processed within approx. 60 seconds, a fault/notice message appears.<br />

Under certain plant conditions, flow values can aggregate in the buffer, for example in the<br />

case <strong>of</strong> prolonged and unwanted fluid backflow. However, this buffer is reset in all<br />

relevant programming adjustments which affect the pulse output.<br />

STANDARD REVERSE<br />

Only negative flow components are totaled. Positive components are not taken into<br />

account.<br />

Factory setting:<br />

STANDARD<br />