<strong>Device</strong> <strong>Functions</strong> <strong>Proline</strong> <strong>Promass</strong> 83<br />
7 Block OUTPUTS<br />
Function description<br />
OUTPUTS PULSE/FREQUENCY OUTPUT (1...2) CONFIGURATION (FREQUENCY)<br />
VALUE F LOW<br />
(4204) ! Note!<br />
This function is not available unless the FREQUENCY setting was selected in the<br />
OPERATION MODE function (4200).<br />
Use this function to assign a variable to the start value frequency (4202).<br />
The value can be higher or lower than the value assigned to the VALUE F HIGH. Positive<br />
and negative values are permissible, depending on the measured variable in question<br />
(e.g. mass flow). You define a measuring range by defining the VALUE F LOW and<br />
VALUE F HIGH values.<br />
User input:<br />
5-digit floating-point number<br />
Factory setting:<br />
0 [kg/h] or 0 [kg/l] or –50 [°C]<br />
! Note!<br />
• For a graphic illustration VALUE F LOW see the VALUE F HIGH function (4205).<br />
• Note that values with different signs cannot be entered for VALUE F LOW<br />
and VALUE F HIGH, if SYMMETRY is the setting selected for the MEASURING<br />
MODE function (4206). In this case, the message “INPUT RANGE EXCEEDED”<br />
appears on the display.<br />
• The appropriate unit is taken from the UNIT MASS FLOW (0400),<br />
UNIT VOLUME FLOW (0402), UNIT CORRECTED VOLUME FLOW (0404), UNIT<br />
DENSITY (0420), UNIT REFERENCE DENSITY (0421) or<br />
UNIT TEMPERATURE (0422) function (see p. 17 to p. 21).<br />
VALUE F HIGH<br />
(4205) ! Note!<br />
This function is not available unless the FREQUENCY setting was selected in the<br />
OPERATION MODE function (4200).<br />
Use this function to assign a variable to the end value frequency (4203).<br />
The value can be higher or lower than the value assigned to the VALUE F LOW. Positive<br />
and negative values are permissible, depending on the measured variable in question<br />
(e.g. mass flow). You define a measuring range by defining the VALUE F LOW and<br />
VALUE F HIGH values.<br />
User input:<br />
5-digit floating-point number<br />
Factory setting:<br />
Depends on nominal diameter [kg/h] or 2 [kg/l] or 200 [°C]<br />
! Note!<br />
Note that values with different signs cannot be entered for VALUE F LOW<br />
and VALUE F HIGH, if SYMMETRY is the setting selected for the MEASURING MODE<br />
function (4206). In this case, the message “INPUT RANGE EXCEEDED” appears on the<br />
display.<br />
125<br />
100<br />
Freq.<br />
0<br />
➀<br />
➁<br />
Q<br />
m = VALUE f min<br />
n = VALUE f max<br />
A0001279<br />
(continued on next page)<br />
Issue/Rev. 0.2 (12/12) MN0M023 • Page 77